ODiSI 6008 8-Channel Optical Distributed Sensor Interrogator (ODiSI 6000 Series)
Eight-channel ODiSI 600x distributed sensing instrument that measures strain and temperature along optical fiber at a 5.2 mm gage pitch.
- Number of channels
- 8
- Gage pitch
- 5.2 mm
- Gages (measurement locations) per meter of sensor
- 192
- Max measurement rate
- 30 Hz
Overview
The ODiSI 6008 is the eight-channel model of the ODiSI 600x line in Luna's ODiSI 6000 Series of optical distributed sensor interrogators. It works on the principle of measuring Rayleigh backscatter in low-cost, unaltered optical fiber. This lets it acquire strain and temperature continuously along each sensor, so a structure or process can be mapped in detail rather than measured at a few discrete points.
The ODiSI 600x models measure with a standard gage pitch of 5.2 mm, and the maximum aggregate measurement rate is 30 Hz. In standard mode with Standard Remote Modules, each channel supports a high-definition sensor up to 20 m long or a continuous fiber grating (CFG) sensor up to 15 m long. With the optional Extended Range Remote Modules, each channel supports sensors up to 100 m long as well as HD-SC strain-compensated temperature sensors.
Software and integration
The system ships with ODiSI application software running on a Linux-based controller. The software is used to identify and configure sensors, view real-time data, log measurements and replay saved files. Data can be streamed in real time over Ethernet using a JSON-based format. Integration features include IEEE 1588 PTP time synchronization, a TTL sync output and start/stop triggering. ODiSI Remote software for Windows and Linux allows remote control and viewing. Optional 3DViz software maps the data onto 3D CAD models or 2D images.
Luna lists the ODiSI 6000 Series as discontinued and names the ODiSI 7100 Series as its replacement.
Key Features
- High-definition distributed strain and temperature sensing based on Rayleigh backscatter
- 8 measurement channels
- Gage pitch of 5.2 mm (192 gages per meter of sensor)
- Up to 100 m of sensing fiber per channel with Extended Range Remote Modules
- Compatible with HD strain and temperature sensors, HD-SC strain-compensated temperature sensors and CFG sensors
- Continuous optical alignment and calibration with every scan, requiring no user intervention
- Real-time streaming of data over Ethernet via TCP-IP
- IEEE 1588 PTP network time synchronization
- JSON-based digital streaming data format
- Sync output TTL pulse
- Start and stop triggering functions
- Passive, corrosion resistant, dielectric sensors
- Long sensor life with no drift or recalibration
Specifications
| Number of channels |
8
ODiSI 600x models are offered with 1, 2, 4 or 8 channels; the ODiSI 6008 is the 8-channel model
|
|---|---|
| Gage pitch |
5.2 mm
Gage pitch is the distance between centerpoints of consecutive gages
|
| Gages (measurement locations) per meter of sensor |
192
At 5.2 mm gage pitch
|
| Max measurement rate | 30 Hz |
| High-Definition (HD) strain & temperature sensor compatibility | Yes |
| Strain-compensated temperature sensor compatibility |
Yes
Requires Extended Range Remote Modules
|
| Continuous Fiber Grating (CFG) sensor compatibility | Yes |
| Data interface | Ethernet |
| Maximum sensor length per channel — standard mode | 20 m |
|---|---|
| Maximum sensor length per channel — extended range |
100 m
Sensor lengths longer than 20 m, as well as HD-SC strain-compensated sensors, require optional Extended Range Remote Modules
|
| Maximum CFG sensor length per channel (standard mode) | 15 m |
| Standoff cable length |
10, 50, 100, 150 or 200 m
Standard lengths of 10, 50 and 100 m; two cables can be combined to form 150 and 200 m
|
| Maximum patch cord length |
Up to 19 m (depending on sensor length)
Patch cord plus active sensor fiber must not exceed 20 m for Standard length mode or 100 m for Extended Range mode
|
| 2.5 m mode |
30 Hz
Rates are aggregate; divide by number of active channels to determine the per-channel rate
|
|---|---|
| 5 m mode |
20 Hz
Aggregate rate
|
| 10 m mode |
12.5 Hz
Aggregate rate
|
| 20 m mode |
6.3 Hz
Aggregate rate
|
| 50 m mode |
2.5 Hz
Aggregate rate
|
| 100 m mode |
1 Hz
Aggregate rate
|
| Strain measurement range | ±15,000 µϵ |
|---|---|
| Resolution | 0.1 µϵ |
| Instrument accuracy | ±1 µϵ |
| System (instrument and sensor) accuracy |
±30 µϵ
Compared to NIST-traceable extensometer measurements; includes errors from ODiSI instrument and Luna strain sensors
|
| Measurement uncertainty |
±2 µϵ
At zero and across the full strain range; twice the standard deviation of 1000 measurements
|
| Dynamic loading rate | 5 Hz |
| Temperature measurement range |
-40 to 200 °C (with standard Luna sensors)
With standard temperature sensors (HD6T and HD6SCT); wider range available with other sensor fiber
|
|---|---|
| Temperature measurement resolution | 0.1 °C |
| Temperature measurement uncertainty |
±0.6 °C
Twice the standard deviation of 1000 measurements
|
| Gage pitch |
6 mm
CFG mode requires Luna CFG sensors
|
|---|---|
| Gages per meter of sensor | 166 |
| Measurement rate — 2.5 m mode |
62.5 Hz
Aggregate rate
|
| Measurement rate — 5 m mode |
40 Hz
Aggregate rate
|
| Measurement rate — 10 m mode |
25 Hz
Aggregate rate
|
| Measurement rate — 20 m mode |
12.5 Hz
Aggregate rate
|
| System (instrument and sensor) accuracy |
±30 µϵ
Preliminary
|
| Measurement uncertainty |
±5 µϵ
Preliminary
|
| Dynamic loading rate | tbd |
| Temperature measurement uncertainty |
±1.0 °C
Preliminary
|
| Gage pitch |
1.04 cm
Requires Luna HD-SC temperature sensors and Extended Range Remote Modules
|
|---|---|
| Measurement rate (all lengths) | 2.5 Hz |
| Resolution | 0.1 °C |
| Measurement uncertainty | ±0.9 °C |
| Strain compensation |
Over an applied strain range of 0 to 1800 µϵ, measurement accuracy is ± 2.8 °C
RMS error with the HD-SC sensor subjected to axial strain; includes effects of the ODiSI interrogator
|
| Class 1 laser | <10 mW |
|---|---|
| Operating temperature range — mainframe and controller | 5 to 40 °C |
| Operating temperature range — standoff cable and remote module | 5 to 60 °C |
| Storage temperature range | 0 to 40 °C |
| Operating and storage relative humidity |
10 to 90 % RH
Non-condensing
|
| Operating altitude | -15 to 3,000 m |
| Dimensions — mainframe (W x D x H) | 34 x 35 x 11 cm |
| Dimensions — laptop controller (W x D x H) | 36 x 24 x 1.7 cm |
| Weight (mainframe and controller) | 17 lb (7.8 kg) |
| Power (mainframe and controller) | 160 W |
Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.
Applications
- High fidelity model verification
- Temperature mapping for batteries
- Improving bonding and welding quality
- Embedded sensors for smart parts
- Strain characterization of new materials and complex structures
- In-situ temperature profiling of critical processes
- Two- and three-dimensional strain field measurement for FE model validation
- Evaluation of multi-material joining
- Detection of small cracks and defects in critical structures
- Strain measurement on and in composite materials
Included
- ODiSI mainframe with 8 channels
- Instrument controller laptop
- ODiSI application software
- One standoff cable for each channel
- One standard length remote module for each channel
Options & Accessories
| Option | Description |
|---|---|
| OP06150 | Extended Range Remote Module — Remote module supporting extended sensor lengths of 50 m and 100 m, in addition to standard sensor lengths and strain-compensated temperature (HD-SC) sensors. |
| OPT06100-R | Rack Mount Controller Option — Rack mount (1U) instrument controller (substituted for laptop). |
| OPT06020 | USB-3106 Analog Output — 16-channel analog output module for ODiSI (voltage or current outputs). |
| OPT06125 | 3Dviz - 3D/2D Visualization Software — Visualization software for mapping real-time or saved ODiSI data onto 3D models and 2D images. Compatible with Windows OS. |
| OPT06123 | 3Dviz - 2D Visualization Software — Visualization software for mapping real-time or saved ODiSI data onto 2D images. Compatible with Windows OS. |