Site Master MS2089A Handheld Cable and Antenna Analyzer with Spectrum Analyzer and VNA Option
Handheld cable and antenna analyzer to 4 or 6 GHz with an integrated 9 kHz to 4/6 GHz spectrum analyzer and optional 2-port, 1-path VNA.
- Frequency
- 5 kHz to 4 GHz
- Frequency
- 5 kHz to 6 GHz
- Frequency Accuracy
- ≤ ± 0.28 ppm (-10ºC to 55ºC) plus aging
- Aging
- ± 1 ppm per year
Overview
The Anritsu Site Master MS2089A is a handheld instrument that combines cable and antenna analysis with an integrated spectrum analyzer, with an optional vector network analyzer (VNA) and vector voltmeter. It is intended for installing, maintaining and troubleshooting RF communication systems up to 6 GHz in telecommunications, broadcasting, DAS, satellite monitoring and aerospace and defense applications.
The MS2089A is offered in two configurations: a 5 kHz to 4 GHz cable and antenna analyzer with a 9 kHz to 4 GHz spectrum analyzer, or a 5 kHz to 6 GHz cable and antenna analyzer with a 9 kHz to 6 GHz spectrum analyzer. The instrument weighs less than 8.5 lb and uses an IK08-rated 10-inch touch screen.
Cable and antenna analysis
The cable and antenna analyzer measures return loss, cable loss, VSWR, distance-to-fault, 1-port phase and Smith chart, with optional time domain reflectometry and 2-port transmission. Built-in PDF/HTML report generation and on-instrument cable lists support close-out documentation of installations.
Spectrum and signal analysis
Compared with the MS2085A, the MS2089A adds a spectrum analyzer for transmitter quality measurements and interference hunting. Options extend it with real-time spectrum analysis, IQ waveform capture and streaming, interference finding, channel scanning, coverage mapping, EMF measurement, AM/FM modulation measurement and WCDMA, LTE FDD/TDD and 5G NR FR1 signal analysis.
Vector network analysis
The optional 2-port, 1-path VNA covers 5 kHz to 4 or 6 GHz with up to 16001 data points, factory ReadyCal plus mechanical and InstaCal user calibrations, and 16 display formats including Log Mag, VSWR, Phase, Smith Chart, Group Delay and Polar.
Key Features
- Cable and antenna analyzer: 5 kHz to 4 GHz or 6 GHz
- Spectrum analyzer: 9 kHz to 4 GHz or 6 GHz
- Optional vector network analyzer: 5 kHz to 4 GHz or 6 GHz
- Return Loss, Cable Loss, VSWR, Distance-To-Fault, Time Domain Reflectometry, Smith Chart, 1-Port Phase
- 105 dB of dynamic range from 50 kHz to 3 GHz typical
- 350 µs/data point in fast sweep mode
- 10,049 data points
- Built-in PDF/HTML report generation
- Add, edit, import and export cable lists on the instrument
- Real-Time Spectrum Analyzer (RTSA)
- IQ capture/streaming
- Interference Analyzer
- True Power Meter
- AM/FM modulation quality measurements
- LTE FDD/TDD Analyzer
- 5GNR FR1 Analyzer
- Coverage mapping
- 2-port, 1-path VNA measurements with 380 µs/point fast sweep
- 16 VNA display formats including Log Mag, VSWR, Phase, Group Delay, Smith Chart, Polar
- Default ReadyCal plus support for Mechanical and InstaCal user calibrations
- Built-in preamplifier included as standard
- IK08-rated 10-inch touch screen
- Explosive Atmosphere MIL-PRF-28800F Section 4.5.6.3 compliant
- Complies with IP52 when installed in soft carrying case
Specifications
|
Frequency
MS2089A-0804
|
5 kHz to 4 GHz
Option 804
|
|---|---|
|
Frequency
MS2089A-0806
|
5 kHz to 6 GHz
Option 806
|
| Frequency Accuracy |
≤ ± 0.28 ppm (-10ºC to 55ºC) plus aging
MS2089A
|
| Aging | ± 1 ppm per year |
| Frequency Resolution | 1 Hz |
| Output Power, High |
0 dBm
typical
|
| Output Power, Low |
–40 dBm
typical
|
| Interference Immunity, On-Channel | +17 dBm @ > 1.0 MHz from carrier frequency |
| Interference Immunity, On-Frequency | 0 dBm within ±10 kHz of the carrier frequency |
| Corrected Directivity |
> 44 dB
typical, OSL Calibration
|
| Corrected Directivity |
> 40 dB
typical, InstaCal Calibration
|
| Measurement Sweep Speed (Return Loss, Distance-to-Fault) |
< 350 µs/data point
RF immunity low, typical; low dynamic range and fast sweep rate. 700 µs/data point normal sweep rate, typical
|
| Return Loss Measurement Range |
0 dB to 60 dB
Resolution 0.01 dB
|
| VSWR Measurement Range |
1:1 to 65:1
Resolution 0.01
|
| Cable Loss Measurement Range |
0 dB to 30 dB
Resolution 0.01 dB
|
| Distance-to-Fault Horizontal Range | 0 to (Data Points–1) x Fault Resolution, to a maximum of 5000 meters (16404 ft) |
| 1-Port Phase Measurement Range |
–450º to +450º
Resolution 0.01º
|
| Smith Chart Marker Resolution | 0.01 Ω, 50/75 Ω selectable |
| Data Points |
Flex Cal: 2 to 10,049, user defined
OSL Calibration: 10,049, 5025, 2513, 1257, 629, 315, 158, 65, 33, 17, 9, 5, 3 and 2
|
| Limit Line Segments | 42 upper and 42 lower segments maximum |
|
Distance
MS2089A-0003
|
5000 Meters |
|---|---|
|
TDR Ohm Measurement Range
MS2089A-0003
|
0 Ω to 5000 Ω
Resolution 0.01 Ω
|
|
TDR Linear Measurement Range
MS2089A-0003
|
0 U to 500 U
Resolution 0.01 U
|
|
Frequency Range
MS2089A-0021
|
5 kHz to 4 GHz (Option 804), 5 kHz to 6 GHz (Option 806) |
|---|---|
|
Dynamic Range, 50 kHz to 3 GHz
MS2089A-0021
|
90 dB, 105 dB typical
Dynamic range set to high
|
|
Dynamic Range, 3 GHz to 4 GHz
MS2089A-0021
|
80 dB, 95 dB typical
Dynamic range set to high
|
|
Dynamic Range, 4 GHz to 6 GHz
MS2089A-0021
|
70 dB, 85 dB typical
Dynamic range set to high
|
|
Frequency
MS2089A-0904
|
5 kHz to 4 GHz
Option 904
|
|---|---|
|
Frequency
MS2089A-0906
|
5 kHz to 6 GHz
Option 906
|
| Frequency Resolution | 1 Hz |
| Transmission Dynamic Range, 5 kHz to < 2 MHz |
110 dB typical
300 Hz IF Bandwidth, 100 point averaging and maximum Port Power
|
| Transmission Dynamic Range, 2 MHz to 2 GHz |
110 dB, 125 dB typical
300 Hz IF Bandwidth, 100 point averaging and maximum Port Power
|
| Transmission Dynamic Range, >2 GHz to 3 GHz |
105 dB, 120 dB typical
300 Hz IF Bandwidth, 100 point averaging and maximum Port Power
|
| Transmission Dynamic Range, > 3 GHz to 4 GHz |
100 dB, 113 dB typical
300 Hz IF Bandwidth, 100 point averaging and maximum Port Power
|
| Transmission Dynamic Range, >4 GHz to 5 GHz |
95 dB, 110 dB typical
300 Hz IF Bandwidth, 100 point averaging and maximum Port Power
|
| Transmission Dynamic Range, >5 GHz to 6 GHz |
90 dB, 105 dB typical
300 Hz IF Bandwidth, 100 point averaging and maximum Port Power
|
| Sweep Speed, 5 kHz to 6 GHz |
380 μs/point
Typical; IF Bandwidth of 100 kHz and 1001 data points
|
| Noise Floor, 2 MHz to 2 GHz |
–116 dBm
300 Hz, typical
|
| Noise Floor, >5 GHz to 6 GHz |
–97 dBm
300 Hz, typical
|
| Maximum Port Power, 2.5 MHz to 5 GHz |
+9 dBm
Default 0 dBm, minimum –45 dBm
|
| Maximum Port Power, > 5 MHz to 6 GHz |
+8 dBm
Default 0 dBm, minimum –45 dBm; band edge as printed in the datasheet
|
| Corrected Directivity, 2 MHz to 6 GHz |
> 44 dB
Typical; full 1-path, 2-port calibration with OSLN50A-8, OSLNF50A-8, TOSLN50A-8 or TOSLNF50A-8
|
| Corrected Source Match, 2 MHz to 6 GHz |
≥ 30 dB
Typical
|
| Data Points | 2-16001 |
| IF Bandwidth |
10 Hz to 1 MHz
Datasheet introduction (page 2) lists IF Bandwidth selections of 10 Hz to 100 kHz
|
|
Frequency
MS2089A-0704
|
9 kHz to 4 GHz
Option 704; usable to 0 Hz
|
|---|---|
|
Frequency
MS2089A-0706
|
9 kHz to 6 GHz
Option 706; usable to 0 Hz
|
| Tuning Resolution | 1 Hz |
| Span | 10 Hz to max frequency, Zero Span |
| Frequency Reference | Internal, GNSS, External |
| Internal Frequency Reference Accuracy |
±0.28 ppm (-10°C ±55°C) plus aging
Standard TCXO; aging ±1 ppm per year
|
| External Frequency Reference | 10 MHz, -10 dBm to +10 dBm |
| Analysis Bandwidth | 20 MHz (standard), 40 MHz (Option 102) |
|---|---|
| Resolution Bandwidth (RBW) | 1 Hz to 5 MHz, 1 Hz to 10 MHz in zero span (standard), 1 Hz to 20 MHz in zero span (Option 102) |
| Video Bandwidth (VBW) | 0.1 Hz to 5 MHz, 1 Hz to 10 MHz in zero span (standard), 1 Hz to 20 MHz in zero span (Option 102) |
| RBW Selectivity | 4:1 nominal (–60 dB / –3 dB) |
| CISPR Bandwidth |
200 Hz, 9 kHz, and 120 kHz
Resolution bandwidth when using Quasi-Peak marker function
|
| Sweep Points | 10 to 10,001 (1001 in zero span) |
| Sweep Rate (non-zero span) | 32 GHz/s (standard), 45 GHz/s (Option 102) |
| Zero Span Sweep Time |
60 ns to 3600 s
Sweep time accuracy ±2 % in zero span
|
| SSB Phase Noise at 10 kHz offset (1 GHz) | –93 dBc/Hz maximum, –94 dBc/Hz typical |
|---|---|
| SSB Phase Noise at 100 kHz offset (1 GHz) | –95 dBc/Hz maximum, –97 dBc/Hz typical |
| SSB Phase Noise at 1 MHz offset (1 GHz) | –120 dBc/Hz maximum, –123 dBc/Hz typical |
| Residual Spurious |
< –120 dBm
RF input terminated, 0 dB input attenuation, > 20 MHz, preamp On; < –105 dBm with preamp Off
|
| Input-Related Spurious |
< –70 dBc
0 dB attenuation, –30 dBm input
|
| Local-Oscillator Related Spurious |
< –60 dBc
nominal for offsets > 1 MHz
|
| Dynamic Range |
105 dB typical at 1 GHz
2/3 (TOI-DANL) in 1 Hz RBW
|
|---|---|
| Measurement Range | DANL to +30 dBm |
| Display Range | 1 to 15 dB/div in 1 dB steps, ten divisions displayed |
| Reference Level Range | –150 dBm to +30 dBm |
| Attenuator Resolution | 0 to 50 dB, 5 dB steps |
| Maximum Continuous Input |
+30 dBm CW, ±50 VDC
≥ 10 dB attenuation; +23 dBm CW (< 10 dB attenuation); +10 dBm CW (preamp ON)
|
| Damage Level | 5 W (+37 dBm) to 6 GHz |
| Amplitude Accuracy, 9 kHz to 6 GHz |
±1.0 dB maximum, ±0.5 dB typical
20°C to 30°C after 30 minute warm-up; ±2.0 dB maximum from –10°C to 55°C after 60 minute warm-up
|
| DANL |
–167 dBm
typical with preamp On
|
| Third-Order Intercept (TOI), 1 GHz |
+7 dBm
Typical; –20 dBm tones 100 kHz apart, 0 dB input attenuation, preamp Off
|
| Third-Order Intercept (TOI), 6 GHz |
+17 dBm
Typical
|
| Second Harmonic Distortion, > 50 MHz to 3 GHz |
–70 dBc
typical; 0 dB input attenuation, –30 dBm input, preamp Off
|
| VSWR, 9 kHz to 2.0 GHz |
1.5:1 typical
≥ 10 dB input attenuation
|
| VSWR, 2 GHz to 6.0 GHz |
1.8:1 typical
≥ 10 dB input attenuation
|
|
RTSA Bandwidth
MS2089A-0199
|
20 MHz (standard), 40 MHz (Option 102) |
|---|---|
|
FFT Rate
MS2089A-0199
|
527,000 FFT/s (normal resolution), 263,000 FFT/s (high resolution) |
|
Minimum Detectable Signal
MS2089A-0199
|
9 ns |
|
Probability of Intercept, 20 MHz span
MS2089A-0199
|
3.036 µs
Normal density resolution, 3 MHz RBW; 4.929 µs high resolution
|
|
Probability of Intercept, 40 MHz span
MS2089A-0199, MS2089A-0102
|
2.464 µs
Option 102, normal density resolution, 5 MHz RBW; 4.357 µs high resolution
|
|
Persistence
MS2089A-0199
|
Infinite or Variable from 0 to 10 s |
|
Acquisition Time
MS2089A-0199
|
50 ms to 5 s |
|
Maximum Sample Rate
MS2089A-0126
|
50 MHz
Option dependent
|
|---|---|
|
Maximum Signal Bandwidth
MS2089A-0126
|
40 MHz
Standard analysis bandwidth up to 20 MHz, Option 102 up to 40 MHz
|
|
Bit Resolution
MS2089A-0126, MS2089A-0127
|
8, 10, 16, or 32-bit |
|
Total Capture Memory
MS2089A-0126
|
2 GB |
|
USB Maximum Gapless Streaming Bandwidth
MS2089A-0127
|
8/10/16 bit: 40 MHz BW, 50 MSPS sample rate; 32 bit: 25 MHz BW, 30.72 MSPS sample rate
Requires USB 3.0 solid state drive
|
|
Frequency Range
MS2089A-0704-0020, MS2089A-0706-0020
|
9 kHz to 6 GHz |
|---|---|
|
Settable Power Range
MS2089A-0704-0020, MS2089A-0706-0020
|
–45 dBm to +9 dBm
Minimum leveled power -40 dBm < 2.5 MHz
|
|
Step Size
MS2089A-0704-0020, MS2089A-0706-0020
|
0.1 dB nominal |
|
Output Connector
MS2089A-0704-0020, MS2089A-0706-0020
|
Type N female, 50 Ω |
|
Channel Scanner Number of Channels
MS2089A-0027
|
1 to 60 |
|---|---|
|
Channel Scanner Frequency Range
MS2089A-0027
|
9 kHz to 4/6 GHz
MS2089A
|
|
Channel Scanner Measurement Range
MS2089A-0027
|
–160 dBm to +30 dBm |
|
GNSS Supported Satellite Systems
MS2089A-0031
|
GNSS (includes combinations of GPS, GLONASS, Galileo, BeiDou)
Requires GNSS antenna, sold separately
|
|
GNSS High Frequency Accuracy
MS2089A-0031
|
< ±2.5 x 10–8
With GNSS On, 3 minutes after satellite lock in selected mode (GNSS antenna connected)
|
|
Gated Sweep Gate Length
MS2089A-0090
|
1 µs up to 200 ms
Gate source GNSS (GPS), External
|
| Battery Life, Cable and Antenna Analyzer | Up to 9 hours |
|---|---|
| Battery Life, Spectrum Analyzer | Up to 6 hours |
| Weight | less than 8.5 lb |
| Display | IK08 10-inch touch screen |
| Ingress Protection |
IP52
When installed in soft carrying case
|
| Explosive Atmosphere | MIL-PRF-28800F Section 4.5.6.3 compliant |
| EMC Directive |
2014/30/EU
EN 61326-1:2013
|
|---|---|
| Low Voltage Directive | 2014/35/EU |
| RoHS Directive | 2011/65/EU + 2015/863 |
| Electrical Safety | IEC 61010-1:2010+AMD1:2016 |
Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.
Applications
- Cellular network installation and maintenance
- Broadcast transmitter systems
- Distributed antenna systems (DAS)
- Military airframe RF cabling
- Ship communications
- Satellite monitoring
- Interference analysis
- Transmitter quality measurements
- Coverage mapping
Options & Accessories
| Option | Description |
|---|---|
| MS2089A-0804 | Frequency Option 4 GHz Cable and Antenna Analyzer |
| MS2089A-0806 | Frequency Option 6 GHz Cable and Antenna Analyzer |
| MS2089A-0704 | Frequency Option 4 GHz Spectrum Analyzer |
| MS2089A-0706 | Frequency Option 6 GHz Spectrum Analyzer |
| MS2089A-0904 | Frequency Option 4 GHz Vector Network Analyzer |
| MS2089A-0906 | Frequency Option 6 GHz Vector Network Analyzer |
| MS2089A-0002 | Time Domain with Gating |
| MS2089A-0003 | Time Domain Reflection — Also offered as a 90-day time-limited option |
| MS2089A-0006 | Remove Wi-Fi and Bluetooth |
| MS2089A-0007 | Secure Data |
| MS2089A-0015 | Vector Voltmeter |
| MS2089A-0017 | Secure Communication |
| MS2089A-0019 | High Accuracy Power Meter — Requires external USB power sensor, sold separately. Also offered as a 90-day time-limited option |
| MS2089A-0021 | 2-Port Transmission Measurement — Also offered as a 90-day time-limited option |
| MS2089A-0024 | Interference Finder — Option 31 and directional antenna recommended. Also offered as a 90-day time-limited option |
| MS2089A-0027 | Channel Scanner — Also offered as a 90-day time-limited option |
| MS2089A-0031 | GNSS Receiver — Requires GNSS antenna, sold separately. Also offered as a 90-day time-limited option |
| MS2089A-0090 | Gated Sweep — Also offered as a 90-day time-limited option |
| MS2089A-0102 | 40 MHz Analysis Bandwidth — Also offered as a 90-day time-limited option |
| MS2089A-0126 | IQ Waveform Capture — Includes MX280005A IQ Signal Master base feature set. Also offered as a 90-day time-limited option |
| MS2089A-0127 | I/Q Waveform Stream — Includes MX280005A IQ Signal Master base feature set; requires Option 126. Also offered as a 90-day time-limited option |
| MS2089A-0128 | Vector Signal Analysis enabled — Also offered as a 90-day time-limited option |
| MS2089A-0199 | Real Time Spectrum Analysis — Also offered as a 90-day time-limited option |
| MS2089A-0400 | Enable Vision Monitor Software — Also offered as a 90-day time-limited option |
| MS2089A-0407 | Enable Vision High-Speed Port Scanner — Also offered as a 90-day time-limited option |
| MS2089A-0431 | Coverage Mapping — Requires Option 31. Also offered as a 90-day time-limited option |
| MS2089A-0444 | EMF Measurement Option — Requires Anritsu isotropic antenna, sold separately. Also offered as a 90-day time-limited option |
| MS2089A-0509 | AM/FM Modulation Measurement — Also offered as a 90-day time-limited option |
| MS2089A-0871 | WCDMA FDD Measurements — Requires Option 31. Also offered as a 90-day time-limited option |
| MS2089A-0883 | LTE FDD/TDD DEMOD — Requires Option 31. Also offered as a 90-day time-limited option |
| MS2089A-0888 | 5G NR Downlink Measurements — Requires Option 31. Also offered as a 90-day time-limited option |
| MS2089A-0704-0020 | Tracking Generator — For the 4 GHz spectrum analyzer configuration |
| MS2089A-0706-0020 | Tracking Generator — For the 6 GHz spectrum analyzer configuration |
| MS2089A-0704-0097 | Accredited Calibration |
| MS2089A-0704-0098 | Standard Calibration |
| MS2089A-0704-0099 | Premium Calibration |
| MS2089A-0706-0097 | Accredited Calibration |
| MS2089A-0706-0098 | Standard Calibration |
| MS2089A-0706-0099 | Premium Calibration |