VIAVI mORL-A1 MORL-A13456-STD-M100-MFA Single-Mode IL/RL Meter Module, 1310/1490/1550/1625 nm, FC/APC
Four-wavelength single-mode insertion loss and return loss meter module (1310/1490/1550/1625 nm, FC/APC) for the VIAVI MAP-Series PCT system.
- Wavelengths
- 1310, 1490, 1550, 1625 nm
- Fiber Type
- Single-Mode 9 µm core
- Output Connector
- FC/APC
Overview
The VIAVI MORL-A13456-STD-M100-MFA is the four-wavelength, standard (non-bidirectional) single-mode version of the mORL-A1 insertion loss and return loss meter. It provides 1310, 1490, 1550 and 1625 nm sources on 9 µm single-mode fiber with an FC/APC output and installs in a single slot of a VIAVI MAP-Series chassis as the core measurement module of the Passive Component/Connector Test (PCT) solution.
Measurement principle
Insertion loss is measured with the power meter method; launch power monitoring and depolarization technology provide 0.001 dB display resolution. Return loss is measured with time-domain technology, often called mandrel-free, which removes the need for manual terminations during setup and test and also measures DUT length. RL measurements run in parallel with IL using the same optical stimulus. According to VIAVI, the single-mode mORL-A1 delivers 80 dB of RL dynamic range and can measure jumpers as short as 70 cm in as little as 6 seconds for two wavelengths.
PCT software
The PCT application runs natively on the MAP chassis in two versions, PCT Max and PCT Lite, and offers instrument mode, script (DUT-aware) mode with a local results database, and port mapping. A full SCPI command set is available over LAN or GPIB for integration with external automation environments. PCT Control Center is a companion PC application for managing DUT definitions, report templates and serial-number batches.
System configuration
The module is often combined with MAP optical switch modules (mOSW-C1), which expand a single output up to 64 channels, and with mOPM remote power meter heads, including the mOPM-C1RHPM Remote Head Polarity Mapper for simultaneous IL, RL and polarity measurement of multifiber assemblies such as MPO. A bidirectional upgrade license is available for use with a high-directivity mOSW 2x2 switch.
Key Features
- Four sources: 1310, 1490, 1550 and 1625 nm on single-mode fiber
- FC/APC output connector
- Time-domain (mandrel-free) return loss measurement with length measurement
- Power meter method insertion loss with 0.001 dB display resolution
- RL measured in parallel with IL using the same optical stimulus
- Scalable modular MAP-Series platform
- Small footprint - 25% of competing solutions
- Integrated automation via SCPI and no-code scripting
- Database interface and report generation
- MPO and MTP capable with simultaneous polarity measurement
- Growth enabling - 1 to 1,000+ DUT channels
- FastIL insertion-loss-only mode for prescreening
Specifications
| Wavelengths |
1310, 1490, 1550, 1625 nm
4-Wavelength Version
|
|---|---|
| Fiber Type |
Single-Mode 9 µm core
Single fiber
|
| Output Connector | FC/APC |
| Initializing Time | < 4s |
|---|---|
| Averaging Options Per Wavelength |
0.5, 1, 2, 5, 10s
The older MAP-Based IL/RL data sheet lists 2, 5, 10s
|
| Total Time (IL, RL, Polarity) for 2 wavelengths |
23s (12 channels); 44s (24 channels)
With mOPM-C1RHPM Remote Head Polarity Mapper
|
| Display Resolution | 0.001 dB |
|---|---|
| Total IL Uncertainty |
± 0.02 dB
After valid zero loss, total expanded uncertainty (2σ), reconnecting the same connector and OPM adaptor, temperature ±1°C, using internal source; IL uncertainty from launching condition not included
|
| Additional Uncertainties Due to 1xN Switching |
± 0.01 dB
If mOSW-C1 added
|
| Additional Uncertainties Due to Fiber Position in the Integrating Sphere |
± 0.03 dB
24-channel ribbon fiber
|
| Display Resolution | 0.01 dB |
|---|---|
| RL Dynamic Range | 80 dB |
| DUT Length, DUT reflections (both ends) < 40 dB | > 170 cm |
| DUT Length, DUT reflections (both ends) > 40 dB | > 70 cm |
| Return Loss Repeatability |
30 to 65 dB ± 0.1 dB; 65 to 70 dB ± 0.2 dB; 70 to 75 dB ± 0.4 dB; 75 to 80 dB ± 1.5 dB
At 5 s averaging time and 200 m range; 10 measurements with a stable connection of a 3 m patch cord
|
| Return Loss Accuracy |
30 to 70 dB ± 1.0 dB; 70 to 75 dB ± 1.7 dB; 75 to 80 dB ± 3.0 dB
At 5 s averaging time and 200 m range
|
| Recalibration Period | 1 year |
|---|
| Warm-up Time | 20 min |
|---|---|
| Operating Temperature, Humidity | 25 ± 5°C non-condensing humidity |
| Storage Temperature | - 30 to + 60°C |
| Size (W x H x D) | 4.06 x 13.26 x 37.03 cm (1.6 x 5.22 x 14.58 in) |
|---|---|
| Weight (Approximately) | 1.2 kg (2.65 lb) |
| Standards |
CE, CSA/UL/IEC61010-1, and LXI Class C requirements
MAP mORL-A1 modules installed in a MAP Series chassis
|
|---|
Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.
Applications
- IL, RL and length testing of optical connectors and cable assemblies
- Structured-cabling solutions and optical splitters testing
- Automated testing of multifiber assemblies such as MPO
- Verifying continuity and polarity of large multifiber assemblies
- Measuring RL of line cards and receptacle-based transponders
- Single-mode fiber-based device testing
Options & Accessories
| Option | Description |
|---|---|
| MSUP-300A-FIT | MAP-300 fiber connector inspection app — Requires probe |
| MSUP-300A-PCTMAPPING | MAP-300 PCT polarity and port mapping application add-on |
| MSUP-300A-PCTREMDB | MAP-300 PCT remote centralized database connection key |
| MSUP-300A-SBSC | MAP-300 PCT driver for legacy SB/SC series switches |
| MSUP-BIDIUPG | MAP-300 mORL Bi-di upgrade license — Requires high directivity mOSW 2x2 |
| MSUP-300A-PCT-TDR | MAP-300 Enhanced PCT TDR License |
| MSUP-300A-PCTCASCADE | MAP-300 PCT Switch Cascade |
| MSUP-300A-PCTLOGGING | MAP-300 PCT Continuous Mode Logging License |
| MSUP-300A-SEIL | MAP-300 Single Ended Insertion Loss license |
| MSUP-FIT | MAP-200 Super Application Fiber Inspection (FIT) |
| MSUP-PCTMAPPING | MAP-200 Super Application PCT Mapping |
| MSUP-PCTREMDB | MAP-200 CT Remote Database Connection Key |
| MSUP-SBSC | MAP-200 driver for legacy SB/SC series switches |
| MOSW-C111C004B0-MXXX-MYY | Single 1 x 4 switch, bulkheads — XXX = fiber type code, YY = connector type code |
| MOSW-C111C008B0-MXXX-MYY | Single 1 x 8 switch, bulkheads — XXX = fiber type code, YY = connector type code |
| MOSW-C111C012B0-MXXX-MYY | Single 1 x 12 switch, bulkheads (Dual width module) — XXX = fiber type code, YY = connector type code |
| MOSW-C111C024B0-MXXX-MYY | Single 1 x 24 switch, bulkheads (Dual width module) — XXX = fiber type code, YY = connector type code |
| MOPM-C1RH1 | Single channel remote interface cassette |
| MOPM-C1RH2 | Dual channel remote interface cassette |
| MOPM-C1RH4 | Quad channel remote interface cassette |
| MOPM-C1RHPCT | 2mm InGaAs PCT system remote head |
| MOPM-C1RHIP | Integrated PCT system remote head |
| MOPM-C1RHPM | Remote Head Polarity Mapper — Requires MAP-300 FW 13.0+ with PCT and MOPM-C1RH1 base cassette |
| AC500 | FC/PC-FC/PC Universal Connector Adapter — Replacement mating sleeve |
| AC501 | FC/PC-SC/PC Universal Connector Adapter — Replacement mating sleeve |
| AC502 | FC/APC-FC/APC Universal Connector Adapter — Replacement mating sleeve |
| AC503 | FC/APC-SC/APC Universal Connector Adapter — Replacement mating sleeve |