VIAVI mORL-A1 MORL-A11308-STD-M112-MFA Multimode IL/RL Meter 850/1300 nm Dual Output OM3/OM1 FC/APC
MAP-Series multimode IL/RL meter module at 850/1300 nm with dual OM3/OM1 output and FC/APC connectors for passive component testing.
- Wavelengths
- 850, 1300 nm
Overview
The MORL-A11308-STD-M112-MFA is a multimode insertion loss (IL) and return loss (RL) meter module from the VIAVI mORL-A1 family. It operates at 850 nm and 1300 nm, has a dual output for 50 µm (OM3) and 62.5 µm (OM1) fiber with software selection between the two, and uses FC/APC connectors. This is the standard (non-bidirectional) version of the dual-fiber multimode module.
The module is part of the VIAVI Passive Component/Connector Test solution (PCT), a family of modules, software and peripherals for testing insertion loss, return loss, physical length and polarity of optical connectivity products such as connectors, cable assemblies, structured cabling and splitters. It installs in a MAP-Series chassis and can be configured for R&D, production or qualification test environments.
Measurement principle
The multimode module uses the same basic technology and architecture as the single-mode mORL-A1. It has an integrated power meter, and IL is measured with the power meter method. RL is measured with time-domain technology, which avoids mandrel terminations. RL can be measured from 15 to 60 dB, concurrently with IL measurements, in less than 2 s per wavelength. IL performance meets the IEC 61280-4-1 recommendations for mode fill. Besides the laser sources used for both IL and RL, the module includes low-power LED sources. These have lower coherence and no polarization, which reduces speckle-related instability at the power meter. Launch power monitoring gives an IL stability of ±0.02 dB.
Software and system integration
The PCT application runs on the MAP chassis in PCT Max and PCT Lite versions. It offers instrument mode, script (DUT-aware) mode with a results database, and port mapping, and a full set of SCPI remote commands is available over LAN or GPIB. The mORL is often combined with MAP optical switch (mOSW/mISW) and remote optical power meter (mOPM) modules to expand channel count and to verify polarity on multifiber assemblies.
Key Features
- Dual output for 50 µm (OM3) and 62.5 µm (OM1) multimode fiber, software selectable
- 850 nm and 1300 nm sources
- LED or laser IL modes, software selectable
- Multimode IL meets IEC 61280-4-1 Encircled Flux requirements
- Time-domain (mandrel-free) RL measurement in parallel with IL
- Integrated power meter
- Scalable modular platform
- Small footprint - 25% of competing solutions
- Integrated automation via SCPI and no-code scripting
- Database interface and report generation
- MPO and MTP capable with simultaneous polarity measurement
- Growth enabling - 1 to 1,000+ DUT channels
- Port mapping verifies multifiber MPO cassette continuity and polarity in less than 15 seconds
- FC/APC connector interface
Specifications
| Wavelengths |
850, 1300 nm
2-wavelength version, multimode mORL-A1
|
|---|
| Dual Fiber | 50 µm core (OM3) and 62.5 µm core (OM1). Software selectable |
|---|---|
| Connector |
FC/APC
Per ordering description: Dual Output OM3/OM1 MMF FC/APC
|
| Initializing Time | < 4s |
|---|---|
| Averaging Options Per Wavelength |
0.5, 1, 2, 5, 10s
Earlier MAP-based IL/RL data sheet lists 2, 5, 10s
|
| Total Time (IL, RL, Polarity) for 2 wavelengths |
23s (12 channels), 44s (24 channels)
Polarity measurement with mOPM-C1RHPM Remote Head Polarity Mapper
|
| RL measurement time (concurrent with IL) |
less than 2s per wavelength
Multimode module, RL from 15 to 60 dB
|
| Modes | LED or laser (software selectable) |
|---|---|
| Display Resolution | 0.001 dB |
| Total IL Uncertainty |
± 0.05 dB
After valid zero loss, total expanded uncertainty (2σ), reconnecting the same connector and OPM adaptor, temperature ±1°C, using internal source (incl. LED mode); IL uncertainty from launching condition not included
|
| Additional Uncertainties Due to 1xN Switching |
± 0.01 dB
If mOSW-C1 added
|
| Additional Uncertainties Due to Fiber Position in the Integrating Sphere |
± 0.03 dB
24-channel ribbon fiber
|
| IL Stability |
±0.02 dB
Achieved through launch power monitoring
|
| Display Resolution | 0.01 dB |
|---|---|
| RL Measurement Range |
15 to 60 dB
Multimode module
|
| DUT Length, DUT reflections (both ends) < 40 dB | > 170 cm |
| DUT Length, DUT reflections (both ends) > 40 dB | > 70 cm |
| Return Loss Repeatability, 15 to 60 dB |
± 0.2 dB
At 5 s averaging time and 200 m range; 10 measurements with a stable connection of a 3 m patch cord
|
| Return Loss Repeatability, 60 to 70 dB |
± 0.5 dB
At 5 s averaging time and 200 m range; 10 measurements with a stable connection of a 3 m patch cord
|
| Return Loss Accuracy, 15 to 20 dB |
± 1.8 dB
At 5 s averaging time and 200 m range
|
| Return Loss Accuracy, 20 to 60 dB |
± 1.3 dB
At 5 s averaging time and 200 m range
|
| Recalibration Period | 1 year |
| Safety and Interface |
CE, CSA/UL/IEC61010-1, and LXI Class C requirements
MAP mORL-A1 modules installed in a MAP Series chassis
|
|---|---|
| Multimode IL | Meets IEC 61280-4-1 Encircled Flux requirements |
| Warm-up Time | 20 min |
|---|---|
| Operating Temperature, Humidity | 25 ± 5°C non-condensing humidity |
| Storage Temperature | - 30 to + 60°C |
| Size (W x H x D) | 4.06 x 13.26 x 37.03 cm (1.6 x 5.22 x 14.58 in) |
|---|---|
| Weight |
1.2 kg (2.65 lb)
Approximately
|
Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.
Applications
- IL, RL and length testing of optical connectors and cable assemblies
- Structured-cabling solution testing
- Optical splitter testing
- Multimode fiber-based device testing
- Automated testing of multifiber assemblies such as MPO
- Verifying continuity and polarity of large multifiber assemblies
- Measuring RL of line cards and receptacle-based transponders
Options & Accessories
| Option | Description |
|---|---|
| MSUP-300A-FIT | MAP-300 fiber connector inspection app — Requires probe |
| MSUP-300A-PCTMAPPING | MAP-300 PCT polarity and port mapping application add-on |
| MSUP-300A-PCTREMDB | MAP-300 PCT remote centralized database connection key |
| MSUP-300A-SBSC | MAP-300 PCT driver for legacy SB/SC series switches |
| MSUP-BIDIUPG | MAP-300 mORL Bi-di upgrade license — Requires high directivity mOSW 2x2 |
| MSUP-300A-PCT-TDR | MAP-300 Enhanced PCT TDR License |
| MSUP-300A-PCTCASCADE | MAP-300 PCT Switch Cascade |
| MSUP-300A-PCTLOGGING | MAP-300 PCT Continuous Mode Logging License |
| MSUP-300A-SEIL | MAP-300 Single Ended Insertion Loss license |
| MSUP-FIT | MAP-200 Super Application Fiber Inspection (FIT) |
| MSUP-PCTMAPPING | MAP-200 Super Application PCT Mapping |
| MSUP-PCTREMDB | MAP-200 CT Remote Database Connection Key |
| MSUP-SBSC | MAP-200 driver for legacy SB/SC series switches |
| MOSW-C111C004B0-MXXX-MYY | Single 1 x 4 switch, bulkheads — XXX = fiber type code, YY = connector type code |
| MOSW-C111C008B0-MXXX-MYY | Single 1 x 8 switch, bulkheads — XXX = fiber type code, YY = connector type code |
| MOSW-C111C012B0-MXXX-MYY | Single 1 x 12 switch, bulkheads (Dual width module) — XXX = fiber type code, YY = connector type code |
| MOSW-C111C024B0-MXXX-MYY | Single 1 x 24 switch, bulkheads (Dual width module) — XXX = fiber type code, YY = connector type code |
| MOPM-C1RH1 | Single channel remote interface cassette |
| MOPM-C1RH2 | Dual channel remote interface cassette |
| MOPM-C1RH4 | Quad channel remote interface cassette |
| MOPM-C1RHPCT | 2mm InGaAs PCT system remote head |
| MOPM-C1RHIP | Integrated PCT system remote head |
| MOPM-C1RHPM | Remote Head Polarity Mapper — Requires MAP-300 FW 13.0+ with PCT |
| AC500 | FC/PC-FC/PC Universal Connector Adapter — Mating sleeve replacement part |
| AC501 | FC/PC-SC/PC Universal Connector Adapter — Mating sleeve replacement part |
| AC502 | FC/APC-FC/APC Universal Connector Adapter — Mating sleeve replacement part |
| AC503 | FC/APC-SC/APC Universal Connector Adapter — Mating sleeve replacement part |