Luna OBR 6235-1 Portable Single-Channel Optical Backscatter Reflectometer
Portable single-channel OFDR reflectometer that measures distributed return loss and insertion loss over fiber links up to 200 m, or 500 m optional.
- Number of optical ports
- 1 port
- Measurement length modes
- 100 m 200 m 500 m
- Maximum device length
- Up to 200 m, (500 m optional)
- Sampling resolution (two-point)
- 0.20 mm 0.40 mm 1.0 mm
Overview
The Luna OBR 6235-1 is the single-channel model of the OBR 6235, a portable, rugged optical backscatter reflectometer in the OBR 6200 Series. It is intended for testing fiber optic links in data centers and longer networks in aerospace, naval and industrial applications.
The instrument uses optical frequency domain reflectometry (OFDR) to measure distributed return loss (RL) and insertion loss (IL) with sub-millimeter spatial resolution, high precision and high dynamic range. It maps reflection versus length and automatically detects RL reflection events and IL locations that exceed user-defined thresholds.
The OBR 6235 is a battery-powered integrated system with a touchscreen user interface, suited to field maintenance work. Its standard measurement range is up to 200 m. The Extended Length Option (OPT06235) adds a 500 m measurement mode for precision testing of data center interconnects.
The OBR 6235-1 has a single standard FC/APC optical connector. It ships with adapter cables with FC/APC and SC/APC connectors.
Key Features
- Fully portable and rugged OBR with extended length range
- Track and analyze return loss (RL) and insertion loss (IL) versus length
- Sub-millimeter sampling resolution with no dead zones
- Measure optical path latency and length with high precision
- Detect and precisely locate reflective events
- Automatic detection of RL reflection events and IL locations that exceed user-defined thresholds
- Single optical channel with standard FC/APC connector
- Battery-powered with 10.1 in touchscreen user interface
Specifications
| Number of optical ports | 1 port |
|---|---|
| Measurement length modes |
100 m 200 m 500 m
500 m mode with extended length option
|
| Maximum device length | Up to 200 m, (500 m optional) |
| Sampling resolution (two-point) |
0.20 mm 0.40 mm 1.0 mm
For 100 m, 200 m and 500 m modes respectively; distance between two sample points in SMF-28. The web page lists 0.20 mm @ 200 m length; the datasheet gives 0.20 mm for the 100 m mode
|
| Time-of-flight delay accuracy |
± 0.005%
Accuracy guaranteed via internal NIST-traceable HCN gas cell
|
| Wavelength scan range |
4 nm 2 nm 0.8 nm
For 100 m, 200 m and 500 m modes respectively
|
| Center wavelength | 1546.7 nm |
| Measurement time | 10 s |
| RL dynamic range |
70 dB
Range between strongest reflection greater than -60 dB and noise floor
|
|---|---|
| Total range |
0 to -129 dB
Noise floor return loss at half of maximum length
|
| Sensitivity |
-129 dB
Noise floor return loss at half of maximum length
|
| Resolution |
± 0.1 dB
Measured with 1 cm integration width
|
| Accuracy |
± 0.5 dB
Measured with 1 cm integration width
|
| IL dynamic range, in reflection mode |
15 dB
Two way loss before backscatter reaches noise floor and IL measurements are no longer possible
|
|---|---|
| Resolution |
± 0.1 dB
Measured with integration widths of 25 cm, 50 cm and 125 cm for 100 m, 200 m and 500 m modes, respectively
|
| Accuracy |
± 0.2 dB
Measured with integration widths of 25 cm, 50 cm and 125 cm for 100 m, 200 m and 500 m modes, respectively
|
| Optical output power | 4 mW |
|---|---|
| Battery runtime | 3 h |
| Battery charge time | 2 h |
| Touchscreen display | 10.1”, 1280 x 800 resolution |
| Data I/O ports | USB-C, RJ45 Ethernet |
| Optical connector | FC/APC (SC/APC or FC/APC adapter patch cord) |
| Weight | 10.1 lb (4.6 kg) |
| Case size | 13.4 x 8.7 x 2.8 in (34 x 22 x 7 cm) |
| Operating temperature |
-20 to 35 °C
0 to 35 °C charging
|
|---|---|
| Storage temperature | -20 to 60 °C |
| Operating altitude |
0 to 2500 m
0 to 3000 m storage
|
Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.
Applications
- Measurement and verification of optical latency in critical interconnects
- Field troubleshooting of fiber assemblies
- Diagnosis and validation of fiber optic links in data centers
- Locating IL sites, high RL connections and fiber breaks
- Maintenance of avionics, aerospace, naval, industrial and data center networks
- Aerospace and avionics fiber optic test
- Fiber optic link length certification
Included
- OBR 6235-1 single-channel system for measurements up to 200 m
- Adapter cables with FC/APC and SC/APC connectors
- Accessory kit
- Power supply/charger
- Ruggedized shipping case
Options & Accessories
| Option | Description |
|---|---|
| OPT06235 | Extended Length Option — Adds 500 m measurement mode to the OBR 6235 |