Luna 6400 Series Lightwave Component Analyzers (Luna 6415, Luna 6435) OFDR Return and Insertion Loss Analyzers
OFDR-based lightwave component analyzers that measure insertion loss, return loss and length in reflection or transmission with 20 μm sampling resolution.
- Measurement length modes, reflection mode
- 20 m, 50 m; Optional: 100 m, 200 m
- Measurement length modes, reflection mode
- 100 m, 200 m, 500 m
- Measurement length modes, transmission mode
- 40 m, 100 m; Optional: 200 m, 400 m
- Measurement length modes, transmission mode
- 200 m, 400 m, 1000 m
Overview
The Luna 6400 Series consists of the Luna 6415 and Luna 6435 lightwave component analyzers for testing passive optical components and fiber optic networks. The instruments measure and analyze the insertion loss (IL) and return loss (RL) distribution, as well as length, scanning the device under test in either reflection or transmission mode.
Measurement principle
The 6400 Series uses optical frequency domain reflectometry (OFDR) to measure backscattered or transmitted light as a function of distance. A sampling resolution of 20 μm and high sensitivity make the Luna 6415 suited to photonic integrated circuits (PICs), silicon photonics and passive components. The Luna 6435 extends the measurement range to 500 m in reflection for testing networks, amplifiers and other fiber systems.
Single-instrument test
By measuring RL, IL and length or delay in reflection or transmission with one instrument, the 6400 Series is intended to reduce the cost and complexity of test and increase throughput in manufacturing. The instruments are controlled remotely via SCPI commands over TCP/IP.
The Luna 6400 Series has been discontinued by the manufacturer; the replacement product is the Lightwave Component Analyzer LWA 7601-C.
Models in this series
| Model | Description | |
|---|---|---|
| Luna 6415 |
Luna 6415 Lightwave Component Analyzer
OFDR-based lightwave component analyzer measuring insertion loss, return loss and length with 20 μm sampling resolution. |
Details → |
| Luna 6435 |
Luna 6435 Lightwave Component Analyzer (OFDR, 500 m Reflection Range)
OFDR-based lightwave component analyzer measuring return loss, insertion loss and length up to 500 m in reflection or 1000 m in transmission. |
Details → |
Key Features
- Accurate return loss (RL) and insertion loss (IL) analyzer
- Track and analyze return loss (RL) and insertion loss (IL) versus length
- Analyze components in both reflection and transmission
- Spectral analysis of IL and RL
- Automatic RL/IL event detection
- Trace distributed loss over length of optical path, with sampling resolution down to 20 μm
- Detect and precisely locate reflective events and measure path length
- Skew measurement with sub-picosecond resolution
- Measurement length to 500 m (in reflection, with Luna 6435)
- Remote control via SCPI commands
Specifications
|
Measurement length modes, reflection mode
Luna 6415 |
20 m, 50 m; Optional: 100 m, 200 m
100 m and 200 m with extended range option OPT06450 |
|---|---|
|
Measurement length modes, reflection mode
Luna 6435 |
100 m, 200 m, 500 m |
|
Measurement length modes, transmission mode
Luna 6415 |
40 m, 100 m; Optional: 200 m, 400 m
200 m and 400 m with extended range option OPT06450 |
|
Measurement length modes, transmission mode
Luna 6435 |
200 m, 400 m, 1000 m |
|
Sampling resolution (two-point)
Luna 6415 |
20 µm
Distance between two sample points along the length axis in SMF-28 |
|
Sampling resolution (two-point)
Luna 6435 |
20 µm, 80 µm
80 µm in 500 m mode (per product page); distance between two sample points along the length axis in SMF-28 |
| Wavelength accuracy |
± 8 pm
Accuracy guaranteed via internal NIST-traceable HCN gas cell |
| Time-of-flight delay accuracy |
± 0.001 %
Accuracy guaranteed via internal NIST-traceable HCN gas cell |
| Center wavelength |
1546.69 nm
Product page states 1546.7 nm; datasheet value used |
|
Wavelength scan range
Luna 6415 |
± 20 nm |
|
Wavelength scan range
Luna 6435 |
± 20 nm, ± 5 nm
Per measurement length mode as printed in the datasheet |
|
Measurement time (in continuous mode)
Luna 6415 |
0.08 s (20 m), 0.8 s (200 m) |
|
Measurement time (in continuous mode)
Luna 6435 |
0.4 s (100 m), 0.8 s (200 m), 0.5 s
Per measurement length mode as printed in the datasheet |
| Maximum optical power | 5 mW |
| RL dynamic range |
70 dB
Range between strongest reflection greater than -60 dB and noise floor |
|---|---|
| Total range |
0 to -130 dB
Noise floor return loss at half of maximum length |
| Sensitivity |
-135 dB
Noise floor return loss at half of maximum length |
| Resolution |
± 0.1 dB
Measured with 1 cm integration width |
| Accuracy |
± 0.2 dB
Measured with 1 cm integration width |
| IL dynamic range, in transmission mode | 70 dB |
|---|---|
| IL dynamic range, in reflection mode |
15 dB
Two way loss before backscatter reaches noise floor and IL measurements are no longer possible |
| Resolution |
± 0.1 dB
With 10 cm integration width for 20 m, and 40 cm for 500 m mode |
| Accuracy |
± 0.2 dB
With 10 cm integration width for 20 m, and 40 cm for 500 m mode |
| Remote interface | SCPI API over TCP/IP |
|---|---|
| Optical connector type | FC/APC |
Applications
- Manufacturing test of fiber optic assemblies
- PIC and silicon photonic testing
- Fiber optic link length and latency testing
- Spatial RL testing
- Automated IL test and analysis
- PLCs, waveguide devices, AWGs, ROADMs
- Filters, couplers, switches, beam splitters
Options & Accessories
| Option | Description |
|---|---|
| OPT06450 | Extended range option — Extends measurement length of the Luna 6415 to 200 m in reflection (400 m in transmission) |