Anritsu MA24350A microwave CW USB power sensor with V(m) connector, front view
Anritsu / Power Sensors / USB Power Sensors
MA24350A

Microwave CW USB Power Sensor MA24350A, 10 MHz to 50 GHz

USB power sensor for CW average power measurements from 10 MHz to 50 GHz over a 90 dB dynamic range, with no zeroing required for most signals.

Frequency Range
10 MHz to 50 GHz
Video Bandwidth
50 kHz
Datasheet

Overview

The Anritsu MA24350A is a microwave CW USB power sensor that measures average power from 10 MHz to 50 GHz across 90 dB of dynamic range through a V(m) connector. It is the 50 GHz member of the MA243x0A series, which also includes the 33 GHz MA24330A and the 40 GHz MA24340A with K(m) connectors. The sensor uses a single-path diode architecture with digital processing that delivers more than 2,100 continuous or 5,600 buffered power readings per second.

A low-noise design removes the need to zero the sensor before measurements for most applications, and no 1 mW reference calibration is required. Internal and external triggering supports time-based measurements, and a list mode speeds up automated test sequences.

Control and connectivity

The sensor connects to a host through a USB 2.0 Micro-B port and can be controlled from a PC with remote programming commands or with PowerXpert, Anritsu's free software application. It is also compatible with most Anritsu RF and microwave handheld instruments; a high accuracy power meter software option such as Option 19 may be required on the instrument.

Ruggedness and calibration

Damage protection of +26 dBm CW and a removable silicone protective covering add durability for field use. The covering is required for warranted operation. Calibration is NIST traceable and traceable to SI units via national metrology institutes, and Anritsu recommends a 12-month calibration interval.

Key Features

  • Frequency range 10 MHz to 50 GHz with V(m) connector
  • Power measurement range +20 dBm to –70 dBm (90 dB dynamic range)
  • CW average power measurements
  • Fast measurement speed: >2,100 readings/s continuous, >5,600 readings/s buffered
  • High damage protection: +26 dBm CW, +32 dBm peak for pulses under 10 µs
  • No zeroing required for signals above –45 dBm
  • No 1 mW reference calibration required
  • Internal and external triggering for time-based measurements
  • External trigger latching for pulses as narrow as 20 ns
  • List mode for faster automated test sequences
  • NIST traceable calibration, traceable to SI units via national metrology institutes
  • Controlled from a PC via remote commands or the free PowerXpert software
  • Compatible with most Anritsu RF and microwave handheld instruments
  • Removable silicone protective covering for field durability
  • Single-path diode architecture
  • Worldwide calibration and service centers

Specifications

Frequency
Frequency Range 10 MHz to 50 GHz
V(m) connector
Video Bandwidth 50 kHz
Power Measurement
Dynamic Range –70 dBm to +20 dBm
Damage Levels at RF Port +26 dBm, ±20 V DC
+32 dBm peak <10 μs pulse and 10 % duty cycle; minimum
Response
Signal Channel Rise Time 8 μs
Characteristic
Sampling Rate 140 kS/s
VSWR
<50 MHz 1.9:1
50 to 150 MHz 1.17:1
>0.15 to 2 GHz 1.08:1
>2 to 6 GHz 1.16:1
>6 to 18 GHz 1.21:1
>18 to 33 GHz 1.29:1
>33 to 40 GHz 1.44:1
>40 to 50 GHz 1.5:1
Trigger
Source Bus, Continuous, Internal, External
Bus trigger is not available in PowerXpert application
Arm Type (for Internal/External) Auto, Single, Multiple, Standby
Internal Trigger
Dynamic Range –35 dBm to +20 dBm
Level Accuracy ±0.5 dB
Characteristic
Slope Positive or Negative
Delay Range –5 ms to +10 s
Delay Resolution 10 μs
Hysteresis 0 dB to 10 dB
0.1 dB resolution
Trigger Hold Off 0 s to 10 s
0.01 ms resolution
External Trigger
External Trigger Input MCX (female), 5.5 V maximum
Input Impedance 4 kΩ
Nominal
Type TTL/CMOS
Slope Positive or Negative
Delay Range –5 ms to +10 s
Delay Resolution 10 μs
High Level Input Voltage 2.3 V min, 3.0 V max
Low Level Input Voltage 1.3 V min, 1.6 V max
Latency 7.1 μs max
Time delay between the defined edge of the applied trigger and the sensor switching into the triggered state
Trigger Pulse Width 20 ns min
Trigger Repetition Period 7.1 μs min
Trigger Holdoff 0 s to 10 s
0.01 ms resolution
Measurement Uncertainty, Average Power (25 °C to 35 °C)
–70 to <+15 dBm, ≤18 GHz 0.11
k=2, CW measurement after zero operation; includes calibration factor and linearity over temperature, excludes mismatch, zero set and drift, and noise
–70 to <+15 dBm, >18 GHz to 40 GHz 0.13
k=2, CW measurement after zero operation
–70 to <+15 dBm, >40 GHz to 50 GHz 0.19
k=2, CW measurement after zero operation
+15 to +20 dBm, ≤18 GHz 0.14
k=2, CW measurement after zero operation
+15 to +20 dBm, >18 GHz to 40 GHz 0.17
k=2, CW measurement after zero operation
+15 to +20 dBm, >40 GHz to 50 GHz 0.23
k=2, CW measurement after zero operation
Measurement Uncertainty, Average Power (0 °C to 50 °C)
–70 to <+15 dBm, ≤18 GHz 0.14
k=2, CW measurement after zero operation; product page lists 0.15 for 10 MHz to 18 GHz
–70 to <+15 dBm, >18 GHz to 40 GHz 0.17
k=2, CW measurement after zero operation
–70 to <+15 dBm, >40 GHz to 50 GHz 0.25
k=2, CW measurement after zero operation; product page lists 0.17 for 18 GHz to 50 GHz combined
+15 to +20 dBm, ≤18 GHz 0.18
k=2, CW measurement after zero operation
+15 to +20 dBm, >18 GHz to 40 GHz 0.21
k=2, CW measurement after zero operation
+15 to +20 dBm, >40 GHz to 50 GHz 0.29
k=2, CW measurement after zero operation; product page lists 0.21 for 18 GHz to 50 GHz combined
Zero Set and Drift
Zero Set, –70 to –20 dBm 9.68E-11 W (–70.14 dBm)
k=3; one hour warm-up followed by Zero; 256 averages, 40 ms aperture, temperature within ±1 °C; valid at 30 °C ± 5 °C
Zero Set, >–20 to 0 dBm 4.95E-09 W (–53.05 dBm)
k=3; valid at 30 °C ± 5 °C
Zero Set, >0 to +20 dBm 1.56E-08 W (–48.08 dBm)
k=3; valid at 30 °C ± 5 °C
Zero Drift, –70 to –20 dBm 8.90E-11 W (–70.50 dBm)
Two sigma value of the reported power over one hour; valid at 30 °C ± 5 °C
Zero Drift, >–20 to 0 dBm 4.14E-09 W (–53.83 dBm)
Two sigma value over one hour; valid at 30 °C ± 5 °C
Zero Drift, >0 to +20 dBm 1.72E-08 W (–47.65 dBm)
Two sigma value over one hour; valid at 30 °C ± 5 °C
Noise
Noise, –70 to –20 dBm 3.53E-11 W (–74.52 dBm)
Two sigma noise at 10.2 seconds of integration time; k=3; valid at 30 °C ± 5 °C
Noise, >–20 to 0 dBm 6.51E-11 W (–71.86 dBm)
Two sigma noise at 10.2 seconds of integration time; k=3
Noise, >0 to +20 dBm 6.30E-10 W (–62.01 dBm)
Two sigma noise at 10.2 seconds of integration time; k=3
System
Measurand Average power
Measurement Resolution 0.01 dB max via PowerXpert, 0.001 dB max via remote command
Offset Correction –100 dB to +150 dB
Available only through the PowerXpert application; no remote command in the sensor firmware
Averaging Auto, Manual
Averaging Type Moving, Repeat
Number of Averages (Manual) 1 to 65,536
Maximum in Continuous Average mode is 65,536; in Scope mode the maximum equals 16,777,216 divided by the number of data points
Auto Average Resolution 1 dB, 0.1 dB, 0.01 dB
0.001 dB resolution is not available with the PowerXpert application
Auto Average Source Scope Data Point Number: 1 to 16,384
Continuous Average Mode
Duty Cycle Correction 0.01 % to 100 %
Aperture Time 0.01 ms to 1 s
Measurement Speed, Continuous >2,100 readings/s
Minimum aperture, one average; measured at the A end of the USB cable; may vary with host CPU speed and load
Measurement Speed, Buffered >5,600 readings/s
Minimum aperture, one average; may vary with host CPU speed and load
Buffer Size 8192
Scope Mode
Capture Time 0.01 ms to 1 s
Data Points 1 to 16,384
Resolution 0.01 ms max
List Mode
Number of Measurements 1 to 1000
Input Parameters Frequency (GHz), aperture time (ms), averages
PowerXpert PC Requirements
Software Version 3.0 or greater
Processor and RAM Minimum: equivalent to Intel Pentium III with 1 GB RAM or Intel Pentium IV with 512 MB RAM; Recommended: equivalent to Intel Pentium IV with 1 GB RAM
Operating System Microsoft Windows 8, Windows 7, and Windows XP
Hard-Disk Free Space 100 MB minimum
Display Resolution 1024 × 768 minimum
Interface USB 2.0 high speed
General
RF Connector V male
Interface to Host USB 2.0 high speed
USB Micro-B port for connectivity to PC or Anritsu handheld instrument
Current Consumption 410 mA to 450 mA
Characteristic, 20 °C to 30 °C
Size 110 mm x 46 mm x 25.6 mm
Excluding V connector and silicone protective covering
Weight 397 g (0.88 lb)
Warranty 1 year
Calibration Cycle 12 months
Anritsu recommended calibration interval
Warm-Up Time 60 minutes
Condition for all specifications unless otherwise stated
Environmental
Operating Temperature Range 0 °C to 50 °C
Tests were performed per MIL-PRF-28800F (Class 3)
Storage Temperature Range –40 °C to +71 °C
Humidity 45 % relative humidity at 50 °C; 75 % relative humidity at 40 °C; 95 % relative humidity at 30 °C
Non-condensing
Altitude 4600 m
Operational max
Shock 30 gn half-sine, 11 ms duration
Vibration, Sinusoidal 5 Hz to 55 Hz, 3 g max
Vibration, Random 10 Hz to 500 Hz, Power Spectral Density 0.03 g²/Hz
Regulatory Compliance
European Union EMC 2014/30/EU, EN 61326:2013, CISPR 11/EN 55011, IEC/EN 61000-4-2/3/4/5/6/8/11
Low Voltage Directive 2014/35/EU
Safety EN 61010-1:2010
RoHS Directive 2011/65/EU
Australia and New Zealand RCM AS/NZS 4417:2012
South Korea KCC-REM-A21-0004

Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.

Applications

  • General purpose RF and microwave power measurement
  • Aerospace and defense
  • Satellite communications
  • Wireless communications
  • Production test

Included

  • 10585-00021 Quick Start Guide
  • 2000-1605-R 1.5 m BNC(m) to MCX(m) cable
  • 2000-2010-R 1.83 m USB A to Micro-B cable

Options & Accessories

Option Description
MA24350A-098 Option 98: Standard calibration ISO/IEC 17025 and ANSI/NCSL Z540-1
MA24350A-099 Option 99: Premium calibration ISO/IEC 17025 and ANSI/NCSL Z540-1 — Includes test report and uncertainty data
01-201 Calibrated torque wrench for K and V connectors
41V-3 Precision fixed attenuator, DC to 60 GHz, 3 dB, 50 Ω, V(m) to V(f)
41V-6 Precision fixed attenuator, DC to 60 GHz, 6 dB, 50 Ω, V(m) to V(f)
41V-10 Precision fixed attenuator, DC to 60 GHz, 10 dB, 50 Ω, V(m) to V(f)
41V-20 Precision fixed attenuator, DC to 60 GHz, 20 dB, 50 Ω, V(m) to V(f)
41KB-3 Precision fixed attenuator, DC to 26.5 GHz, 3 dB, 50 Ω, K(m) to K(f)
41KB-6 Precision fixed attenuator, DC to 26.5 GHz, 6 dB, 50 Ω, K(m) to K(f)
41KB-10 Precision fixed attenuator, DC to 26.5 GHz, 10 dB, 50 Ω, K(m) to K(f)
41KB-20 Precision fixed attenuator, DC to 26.5 GHz, 20 dB, 50 Ω, K(m) to K(f)
41KC-3 Precision fixed attenuator, DC to 40 GHz, 3 dB, 50 Ω, K(m) to K(f)
41KC-6 Precision fixed attenuator, DC to 40 GHz, 6 dB, 50 Ω, K(m) to K(f)
41KC-10 Precision fixed attenuator, DC to 40 GHz, 10 dB, 50 Ω, K(m) to K(f)
41KC-20 Precision fixed attenuator, DC to 40 GHz, 20 dB, 50 Ω, K(m) to K(f)
43KB-3 Precision fixed attenuator, DC to 26.5 GHz, 3 dB, 50 Ω, K(m) to K(f)
43KB-6 Precision fixed attenuator, DC to 26.5 GHz, 6 dB, 50 Ω, K(m) to K(f)
43KB-10 Precision fixed attenuator, DC to 26.5 GHz, 10 dB, 50 Ω, K(m) to K(f)
43KB-20 Precision fixed attenuator, DC to 26.5 GHz, 20 dB, 50 Ω, K(m) to K(f)
43KC-3 Precision fixed attenuator, DC to 40 GHz, 3 dB, 50 Ω, K(m) to K(f)
43KC-6 Precision fixed attenuator, DC to 40 GHz, 6 dB, 50 Ω, K(m) to K(f)
43KC-10 Precision fixed attenuator, DC to 40 GHz, 10 dB, 50 Ω, K(m) to K(f)
43KC-20 Precision fixed attenuator, DC to 40 GHz, 20 dB, 50 Ω, K(m) to K(f)
33VFVF50C Precision coaxial adapter, DC to 70 GHz, 50 Ω, V(f) to V(f)
33VVF50C Precision coaxial adapter, DC to 70 GHz, 50 Ω, V(m) to V(f)
33KFKF50B Precision coaxial adapter, DC to 40 GHz, 50 Ω, K(f) to K(f)
33KKF50B Precision coaxial adapter, DC to 40 GHz, 50 Ω, K(m) to K(f)
34NKF50 Precision coaxial adapter, DC to 18 GHz, 50 Ω, N(m) to K(f)
34NFKF50 Precision coaxial adapter, DC to 18 GHz, 50 Ω, N(f) to K(f)