Microwave CW USB Power Sensor MA24350A, 10 MHz to 50 GHz
USB power sensor for CW average power measurements from 10 MHz to 50 GHz over a 90 dB dynamic range, with no zeroing required for most signals.
- Frequency Range
- 10 MHz to 50 GHz
- Video Bandwidth
- 50 kHz
Overview
The Anritsu MA24350A is a microwave CW USB power sensor that measures average power from 10 MHz to 50 GHz across 90 dB of dynamic range through a V(m) connector. It is the 50 GHz member of the MA243x0A series, which also includes the 33 GHz MA24330A and the 40 GHz MA24340A with K(m) connectors. The sensor uses a single-path diode architecture with digital processing that delivers more than 2,100 continuous or 5,600 buffered power readings per second.
A low-noise design removes the need to zero the sensor before measurements for most applications, and no 1 mW reference calibration is required. Internal and external triggering supports time-based measurements, and a list mode speeds up automated test sequences.
Control and connectivity
The sensor connects to a host through a USB 2.0 Micro-B port and can be controlled from a PC with remote programming commands or with PowerXpert, Anritsu's free software application. It is also compatible with most Anritsu RF and microwave handheld instruments; a high accuracy power meter software option such as Option 19 may be required on the instrument.
Ruggedness and calibration
Damage protection of +26 dBm CW and a removable silicone protective covering add durability for field use. The covering is required for warranted operation. Calibration is NIST traceable and traceable to SI units via national metrology institutes, and Anritsu recommends a 12-month calibration interval.
Key Features
- Frequency range 10 MHz to 50 GHz with V(m) connector
- Power measurement range +20 dBm to –70 dBm (90 dB dynamic range)
- CW average power measurements
- Fast measurement speed: >2,100 readings/s continuous, >5,600 readings/s buffered
- High damage protection: +26 dBm CW, +32 dBm peak for pulses under 10 µs
- No zeroing required for signals above –45 dBm
- No 1 mW reference calibration required
- Internal and external triggering for time-based measurements
- External trigger latching for pulses as narrow as 20 ns
- List mode for faster automated test sequences
- NIST traceable calibration, traceable to SI units via national metrology institutes
- Controlled from a PC via remote commands or the free PowerXpert software
- Compatible with most Anritsu RF and microwave handheld instruments
- Removable silicone protective covering for field durability
- Single-path diode architecture
- Worldwide calibration and service centers
Specifications
| Frequency Range |
10 MHz to 50 GHz
V(m) connector
|
|---|---|
| Video Bandwidth | 50 kHz |
| Dynamic Range | –70 dBm to +20 dBm |
|---|---|
| Damage Levels at RF Port |
+26 dBm, ±20 V DC
+32 dBm peak <10 μs pulse and 10 % duty cycle; minimum
|
| Signal Channel Rise Time |
8 μs
Characteristic
|
|---|---|
| Sampling Rate | 140 kS/s |
| <50 MHz | 1.9:1 |
|---|---|
| 50 to 150 MHz | 1.17:1 |
| >0.15 to 2 GHz | 1.08:1 |
| >2 to 6 GHz | 1.16:1 |
| >6 to 18 GHz | 1.21:1 |
| >18 to 33 GHz | 1.29:1 |
| >33 to 40 GHz | 1.44:1 |
| >40 to 50 GHz | 1.5:1 |
| Source |
Bus, Continuous, Internal, External
Bus trigger is not available in PowerXpert application
|
|---|---|
| Arm Type (for Internal/External) | Auto, Single, Multiple, Standby |
| Dynamic Range | –35 dBm to +20 dBm |
|---|---|
| Level Accuracy |
±0.5 dB
Characteristic
|
| Slope | Positive or Negative |
| Delay Range | –5 ms to +10 s |
| Delay Resolution | 10 μs |
| Hysteresis |
0 dB to 10 dB
0.1 dB resolution
|
| Trigger Hold Off |
0 s to 10 s
0.01 ms resolution
|
| External Trigger Input | MCX (female), 5.5 V maximum |
|---|---|
| Input Impedance |
4 kΩ
Nominal
|
| Type | TTL/CMOS |
| Slope | Positive or Negative |
| Delay Range | –5 ms to +10 s |
| Delay Resolution | 10 μs |
| High Level Input Voltage | 2.3 V min, 3.0 V max |
| Low Level Input Voltage | 1.3 V min, 1.6 V max |
| Latency |
7.1 μs max
Time delay between the defined edge of the applied trigger and the sensor switching into the triggered state
|
| Trigger Pulse Width | 20 ns min |
| Trigger Repetition Period | 7.1 μs min |
| Trigger Holdoff |
0 s to 10 s
0.01 ms resolution
|
| –70 to <+15 dBm, ≤18 GHz |
0.11
k=2, CW measurement after zero operation; includes calibration factor and linearity over temperature, excludes mismatch, zero set and drift, and noise
|
|---|---|
| –70 to <+15 dBm, >18 GHz to 40 GHz |
0.13
k=2, CW measurement after zero operation
|
| –70 to <+15 dBm, >40 GHz to 50 GHz |
0.19
k=2, CW measurement after zero operation
|
| +15 to +20 dBm, ≤18 GHz |
0.14
k=2, CW measurement after zero operation
|
| +15 to +20 dBm, >18 GHz to 40 GHz |
0.17
k=2, CW measurement after zero operation
|
| +15 to +20 dBm, >40 GHz to 50 GHz |
0.23
k=2, CW measurement after zero operation
|
| –70 to <+15 dBm, ≤18 GHz |
0.14
k=2, CW measurement after zero operation; product page lists 0.15 for 10 MHz to 18 GHz
|
|---|---|
| –70 to <+15 dBm, >18 GHz to 40 GHz |
0.17
k=2, CW measurement after zero operation
|
| –70 to <+15 dBm, >40 GHz to 50 GHz |
0.25
k=2, CW measurement after zero operation; product page lists 0.17 for 18 GHz to 50 GHz combined
|
| +15 to +20 dBm, ≤18 GHz |
0.18
k=2, CW measurement after zero operation
|
| +15 to +20 dBm, >18 GHz to 40 GHz |
0.21
k=2, CW measurement after zero operation
|
| +15 to +20 dBm, >40 GHz to 50 GHz |
0.29
k=2, CW measurement after zero operation; product page lists 0.21 for 18 GHz to 50 GHz combined
|
| Zero Set, –70 to –20 dBm |
9.68E-11 W (–70.14 dBm)
k=3; one hour warm-up followed by Zero; 256 averages, 40 ms aperture, temperature within ±1 °C; valid at 30 °C ± 5 °C
|
|---|---|
| Zero Set, >–20 to 0 dBm |
4.95E-09 W (–53.05 dBm)
k=3; valid at 30 °C ± 5 °C
|
| Zero Set, >0 to +20 dBm |
1.56E-08 W (–48.08 dBm)
k=3; valid at 30 °C ± 5 °C
|
| Zero Drift, –70 to –20 dBm |
8.90E-11 W (–70.50 dBm)
Two sigma value of the reported power over one hour; valid at 30 °C ± 5 °C
|
| Zero Drift, >–20 to 0 dBm |
4.14E-09 W (–53.83 dBm)
Two sigma value over one hour; valid at 30 °C ± 5 °C
|
| Zero Drift, >0 to +20 dBm |
1.72E-08 W (–47.65 dBm)
Two sigma value over one hour; valid at 30 °C ± 5 °C
|
| Noise, –70 to –20 dBm |
3.53E-11 W (–74.52 dBm)
Two sigma noise at 10.2 seconds of integration time; k=3; valid at 30 °C ± 5 °C
|
|---|---|
| Noise, >–20 to 0 dBm |
6.51E-11 W (–71.86 dBm)
Two sigma noise at 10.2 seconds of integration time; k=3
|
| Noise, >0 to +20 dBm |
6.30E-10 W (–62.01 dBm)
Two sigma noise at 10.2 seconds of integration time; k=3
|
| Measurand | Average power |
|---|---|
| Measurement Resolution | 0.01 dB max via PowerXpert, 0.001 dB max via remote command |
| Offset Correction |
–100 dB to +150 dB
Available only through the PowerXpert application; no remote command in the sensor firmware
|
| Averaging | Auto, Manual |
| Averaging Type | Moving, Repeat |
| Number of Averages (Manual) |
1 to 65,536
Maximum in Continuous Average mode is 65,536; in Scope mode the maximum equals 16,777,216 divided by the number of data points
|
| Auto Average Resolution |
1 dB, 0.1 dB, 0.01 dB
0.001 dB resolution is not available with the PowerXpert application
|
| Auto Average Source | Scope Data Point Number: 1 to 16,384 |
| Duty Cycle Correction | 0.01 % to 100 % |
|---|---|
| Aperture Time | 0.01 ms to 1 s |
| Measurement Speed, Continuous |
>2,100 readings/s
Minimum aperture, one average; measured at the A end of the USB cable; may vary with host CPU speed and load
|
| Measurement Speed, Buffered |
>5,600 readings/s
Minimum aperture, one average; may vary with host CPU speed and load
|
| Buffer Size | 8192 |
| Capture Time | 0.01 ms to 1 s |
|---|---|
| Data Points | 1 to 16,384 |
| Resolution | 0.01 ms max |
| Number of Measurements | 1 to 1000 |
|---|---|
| Input Parameters | Frequency (GHz), aperture time (ms), averages |
| Software Version | 3.0 or greater |
|---|---|
| Processor and RAM | Minimum: equivalent to Intel Pentium III with 1 GB RAM or Intel Pentium IV with 512 MB RAM; Recommended: equivalent to Intel Pentium IV with 1 GB RAM |
| Operating System | Microsoft Windows 8, Windows 7, and Windows XP |
| Hard-Disk Free Space | 100 MB minimum |
| Display Resolution | 1024 × 768 minimum |
| Interface | USB 2.0 high speed |
| RF Connector | V male |
|---|---|
| Interface to Host |
USB 2.0 high speed
USB Micro-B port for connectivity to PC or Anritsu handheld instrument
|
| Current Consumption |
410 mA to 450 mA
Characteristic, 20 °C to 30 °C
|
| Size |
110 mm x 46 mm x 25.6 mm
Excluding V connector and silicone protective covering
|
| Weight | 397 g (0.88 lb) |
| Warranty | 1 year |
| Calibration Cycle |
12 months
Anritsu recommended calibration interval
|
| Warm-Up Time |
60 minutes
Condition for all specifications unless otherwise stated
|
| Operating Temperature Range |
0 °C to 50 °C
Tests were performed per MIL-PRF-28800F (Class 3)
|
|---|---|
| Storage Temperature Range | –40 °C to +71 °C |
| Humidity |
45 % relative humidity at 50 °C; 75 % relative humidity at 40 °C; 95 % relative humidity at 30 °C
Non-condensing
|
| Altitude |
4600 m
Operational max
|
| Shock | 30 gn half-sine, 11 ms duration |
| Vibration, Sinusoidal | 5 Hz to 55 Hz, 3 g max |
| Vibration, Random | 10 Hz to 500 Hz, Power Spectral Density 0.03 g²/Hz |
| European Union EMC | 2014/30/EU, EN 61326:2013, CISPR 11/EN 55011, IEC/EN 61000-4-2/3/4/5/6/8/11 |
|---|---|
| Low Voltage Directive | 2014/35/EU |
| Safety | EN 61010-1:2010 |
| RoHS Directive | 2011/65/EU |
| Australia and New Zealand | RCM AS/NZS 4417:2012 |
| South Korea | KCC-REM-A21-0004 |
Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.
Applications
- General purpose RF and microwave power measurement
- Aerospace and defense
- Satellite communications
- Wireless communications
- Production test
Included
- 10585-00021 Quick Start Guide
- 2000-1605-R 1.5 m BNC(m) to MCX(m) cable
- 2000-2010-R 1.83 m USB A to Micro-B cable
Options & Accessories
| Option | Description |
|---|---|
| MA24350A-098 | Option 98: Standard calibration ISO/IEC 17025 and ANSI/NCSL Z540-1 |
| MA24350A-099 | Option 99: Premium calibration ISO/IEC 17025 and ANSI/NCSL Z540-1 — Includes test report and uncertainty data |
| 01-201 | Calibrated torque wrench for K and V connectors |
| 41V-3 | Precision fixed attenuator, DC to 60 GHz, 3 dB, 50 Ω, V(m) to V(f) |
| 41V-6 | Precision fixed attenuator, DC to 60 GHz, 6 dB, 50 Ω, V(m) to V(f) |
| 41V-10 | Precision fixed attenuator, DC to 60 GHz, 10 dB, 50 Ω, V(m) to V(f) |
| 41V-20 | Precision fixed attenuator, DC to 60 GHz, 20 dB, 50 Ω, V(m) to V(f) |
| 41KB-3 | Precision fixed attenuator, DC to 26.5 GHz, 3 dB, 50 Ω, K(m) to K(f) |
| 41KB-6 | Precision fixed attenuator, DC to 26.5 GHz, 6 dB, 50 Ω, K(m) to K(f) |
| 41KB-10 | Precision fixed attenuator, DC to 26.5 GHz, 10 dB, 50 Ω, K(m) to K(f) |
| 41KB-20 | Precision fixed attenuator, DC to 26.5 GHz, 20 dB, 50 Ω, K(m) to K(f) |
| 41KC-3 | Precision fixed attenuator, DC to 40 GHz, 3 dB, 50 Ω, K(m) to K(f) |
| 41KC-6 | Precision fixed attenuator, DC to 40 GHz, 6 dB, 50 Ω, K(m) to K(f) |
| 41KC-10 | Precision fixed attenuator, DC to 40 GHz, 10 dB, 50 Ω, K(m) to K(f) |
| 41KC-20 | Precision fixed attenuator, DC to 40 GHz, 20 dB, 50 Ω, K(m) to K(f) |
| 43KB-3 | Precision fixed attenuator, DC to 26.5 GHz, 3 dB, 50 Ω, K(m) to K(f) |
| 43KB-6 | Precision fixed attenuator, DC to 26.5 GHz, 6 dB, 50 Ω, K(m) to K(f) |
| 43KB-10 | Precision fixed attenuator, DC to 26.5 GHz, 10 dB, 50 Ω, K(m) to K(f) |
| 43KB-20 | Precision fixed attenuator, DC to 26.5 GHz, 20 dB, 50 Ω, K(m) to K(f) |
| 43KC-3 | Precision fixed attenuator, DC to 40 GHz, 3 dB, 50 Ω, K(m) to K(f) |
| 43KC-6 | Precision fixed attenuator, DC to 40 GHz, 6 dB, 50 Ω, K(m) to K(f) |
| 43KC-10 | Precision fixed attenuator, DC to 40 GHz, 10 dB, 50 Ω, K(m) to K(f) |
| 43KC-20 | Precision fixed attenuator, DC to 40 GHz, 20 dB, 50 Ω, K(m) to K(f) |
| 33VFVF50C | Precision coaxial adapter, DC to 70 GHz, 50 Ω, V(f) to V(f) |
| 33VVF50C | Precision coaxial adapter, DC to 70 GHz, 50 Ω, V(m) to V(f) |
| 33KFKF50B | Precision coaxial adapter, DC to 40 GHz, 50 Ω, K(f) to K(f) |
| 33KKF50B | Precision coaxial adapter, DC to 40 GHz, 50 Ω, K(m) to K(f) |
| 34NKF50 | Precision coaxial adapter, DC to 18 GHz, 50 Ω, N(m) to K(f) |
| 34NFKF50 | Precision coaxial adapter, DC to 18 GHz, 50 Ω, N(f) to K(f) |