VIAVI OCETS Plus Optical Component Environmental Test System
Automated test system that tracks optical component insertion and return loss inside a temperature/humidity chamber during environmental stress testing.
- Fiber type
- 9/125 μm fiber, standard 3 mm jacket
- Fiber type
- 50/125 μm fiber, standard 3 mm jacket
- SC switch pigtail lengths (to DUTs)
- 5 m per switch (10 m total between the two switches)
- Insertion Loss (IL) dynamic range
- >65 dB
Overview
The VIAVI OCETS Plus (Optical Component Environmental Test System) is an integrated, automated test facility for checking the long-term reliability of optical components. It measures components while they are under environmental stress in a temperature/humidity chamber. It is the third generation of the classic OCETS. VIAVI improved its hardware specifications and software algorithms to meet current requirements for optical component qualification testing, such as those of the Verizon FOC program.
Qualification standards require a representative set of sample devices to go through a program of environmental stresses. These include dwelling at high and low temperatures at target humidity levels and cycling between temperatures. Device characteristics must be measured before and after each stage, and sometimes continuously or at intervals during a stage. OCETS Plus makes these optical measurements without removing the devices from the environmental chamber.
Measurement hardware
The core of the system is a pair of custom-grade programmable 1xN optical switches based on the MAP-200 MISW-C1 Optical Switch. These switches are specified for higher insertion loss repeatability and background return loss than analogue-grade switches. They use beam traps to capture stray light, which optimizes them for return-loss stability. The platform also includes light sources, a source switch, directional switches, a polarization controller, high-directivity couplers and an optical power meter. Single-mode systems use individually filtered Super Luminescent Diodes (SLEDs), which give high power and a short coherence length for long-term stability.
Software
The system comes with two software packages. EasyOCETS2017 is the main control and data-collection interface, and EasyViewer 2017 is used to view and extract data. Data is stored in an SQL database that can be hosted anywhere on the network. Test files, DUT definitions and system configuration files are kept in the database and can be shared between systems. The test executor and scheduler can run multiple tests at the same time. Simulation and continuity-check modes, a DUT library editor and auto-channel-connection algorithms help with setting up tests.
Key Features
- High insertion loss (IL) and return loss (RL) repeatability
- High Return Loss option (HiRL) monitors RL up to 70 dB
- Supplied with EasyOCETS software
- Single-mode and multimode systems
- Full bidirectional testing
- Up to 320 device channels (640 ports)
- Custom-grade programmable switches based on the MAP-200 MISW-C1 Optical Switch
- SQL database that can be remotely hosted and shared between systems
- Simulation and continuity-check modes
Specifications
|
Fiber type
Single-Mode (SM)
|
9/125 μm fiber, standard 3 mm jacket |
|---|---|
|
Fiber type
Multimode (MM)
|
50/125 μm fiber, standard 3 mm jacket |
| SC switch pigtail lengths (to DUTs) | 5 m per switch (10 m total between the two switches) |
|
Insertion Loss (IL) dynamic range
Single-Mode (SM)
|
>65 dB |
|
Insertion Loss (IL) dynamic range
Multimode (MM)
|
>50 dB |
|
Insertion Loss (IL) repeatability over 100 hours
Single-Mode (SM)
|
±0.04 dB for IL <50 dB |
|
Insertion Loss (IL) repeatability over 100 hours
Multimode (MM)
|
±0.04 dB for IL <35 dB |
|
Return Loss dynamic range
Single-Mode (SM)
|
>70 dB |
|
Return Loss dynamic range
Multimode (MM)
|
>35 dB |
|
Return Loss (RL) repeatability over 100 hours
Single-Mode (SM)
|
±0.4 dB up to 55 dB |
|
Return Loss (RL) repeatability over 100 hours
Single-Mode (SM), HiRL
|
±1 dB up to 65 dB; ±3 dB up to 70 dB
With HiRL option
|
|
Return Loss (RL) repeatability over 100 hours
Multimode (MM)
|
±0.4 dB up to 30 dB |
|
Measurement timing
Single-Mode (SM)
|
IL, RL, HiRL <2.5 s
Averaged over 60 consecutive measurements, not including reference or saving to database
|
|
Measurement timing
Multimode (MM)
|
IL, RL <2 s
Averaged over 60 consecutive measurements, not including reference or saving to database
|
|
Sources available
Single-Mode (SM)
|
1310, 1490, 1550, 1625 ±10 nm SLED |
|
Sources available
Multimode (MM)
|
850, 1300 ±20 nm LED
Multimode launch conditions meet the requirements of IEC 61280-4-1 Ed. 2.0
|
|
Source power stability at 23°C
Single-Mode (SM)
|
±0.01 dB for 20 minutes
Measured at full power, constant current mode in a controlled environment of 23±1°C
|
|
Source power stability at 23°C
Multimode (MM)
|
±0.01 dB for 15 minutes
Measured at full power, constant current mode in a controlled environment of 23±1°C
|
| Number of channels |
55, 105, 160, 210, or 320 input and output channels
e.g. 55 channels means 55 input and 55 output channels. The web page states up to 210 device channels (420 ports); the datasheet lists up to 320.
|
|---|---|
| Number of reference channels |
1 IL path, 1 RL path (per DUT switch)
The RL reference used by the EasyOCETS2017 algorithms is a 0 dB reflector; users may add RL references to other ports, and as many ports as required can be used for IL references
|
| Equipment warm-up time | 4 hours, can be left on indefinitely with no adverse side effects |
| Input voltage | 100V to 240V AC, 50/60 Hz auto-switching |
| Power consumption (includes computer) | 55 to 160 channels: 750 VA; 210 to 320 channels: 950 VA |
| Computer control | Minimum configuration: Intel Core i3 CPU, 8GB RAM, 500GB HDD, 2 x 10/100G Ethernet interface, USB 2.0 port |
| Mechanical configuration | All equipment, except computer, is installed in a single bay, 32U, 19-inch rack with removable covers and door; cabinet includes top-mounted fans, casters and levelers; rear door access to MAP cassettes |
|---|---|
| Dimensions (W x H x D) | 22 x 72 x 26 in (56 x 183 x 66 cm) |
| Weight | 55 to 160 channels: 190 kg; 320 channels: 220 kg |
| Operating humidity |
0 to 80% RH range
Maximum variation range during a test: 15% RH
|
|---|---|
| Operating temperature |
15 to 30°C range
Maximum variation within range during a test: 3°C
|
| Compliance | Complies to CE requirements. Switch and MAP based products comply to UL3101.1 and CAN/CSA-C22.2 No. 1010.1 |
|---|---|
| Laser classification |
MAP lasers are Class 1 except for 850 nm version which is Class 1M
Classified per IEC 60825-1(2002); comply with 21CFR1040.10 except deviations per Laser Notice No. 50, July 2001
|
| DUT switch connection | FC/APC bulkhead connectors on the DUT switches, or 10-meter unterminated pigtails |
|---|---|
|
High Return Loss (HiRL) option
Single-Mode (SM)
|
Single Mode systems only; only available for systems supplied with 10-meter unterminated pigtails |
Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.
Applications
- Unattended long-term monitoring of optical component IL and RL
- Telcordia GR-326-CORE, GR-910-CORE, GR-1435-CORE, GR-1209-CORE and GR-2866-CORE testing
- Verizon FOC qualification of jumpers, cables and passive splitters
- Optical component qualification testing under environmental stress
Included
- EasyOCETS2017 software
- EasyViewer 2017 software
Options & Accessories
| Option | Description |
|---|---|
| MOCETS-A4055S0-M100-MXX | Standard single-mode, IL and standard RL, 55 channels |
| MOCETS-A4105S0-M100-MXX | Standard single-mode, IL and standard RL, 105 channels |
| MOCETS-A4160S0-M100-MXX | Standard single-mode, IL and standard RL, 160 channels |
| MOCETS-A4210S0-M100-MXX | Standard single-mode, IL and standard RL, 210 channels |
| MOCETS-A4320S0-M100-MXX | Standard single-mode, IL and standard RL, 320 channels |
| MOCETS-A4055S0-M101-MXX | 50um multimode EF compliant, 55 channels |
| MOCETS-A4105S0-M101-MXX | 50um multimode EF compliant, 105 channels |
| MOCETS-A4160S0-M101-MXX | 50um multimode EF compliant, 160 channels |
| MOCETS-A4210S0-M101-MXX | 50um multimode EF compliant, 210 channels |
| MOCETS-A4320S0-M101-MXX | 50um multimode EF compliant, 320 channels |
| MOCETS-A4055H0-M100-MNC | Standard single-mode, IL and High RL, 55 channels |
| MOCETS-A4105H0-M100-MNC | Standard single-mode, IL and High RL, 105 channels |
| MOCETS-A4160H0-M100-MNC | Standard single-mode, IL and High RL, 160 channels |
| MOCETS-A4210H0-M100-MNC | Standard single-mode, IL and High RL, 210 channels |
| MOCETS-A4320H0-M100-MNC | Standard single-mode, IL and High RL, 320 channels |
| MFA | FC/APC connector type (MXX code) |
| MNC | No connectors (MXX code) |