Luna PDL-201 Polarization Dependent Loss Multimeter front panel
Luna / Optical Component Test / Instruments and Meters
PDL-201

PolaChex PDL-201 Polarization Dependent Loss (PDL) Multimeter

Benchtop multimeter that measures polarization dependent loss, insertion loss and optical power from 1260 to 1620 nm in 30 ms.

Wavelength Range
1260 to 1620 nm
Resolution
0.01 dB
PDL Accuracy
± (0.01 + 5% of PDL) (dB)
PDL Repeatability
± (0.005 + 2% of PDL) (dB)
Datasheet

Overview

The Luna PDL-201 (PolaChex) measures the polarization dependent loss (PDL), insertion loss (IL) and optical power of a device under test at the same time, taking 30 ms per measurement. It uses a patented maximum and minimum search method that complies with TIA/EIA-455-198.

PDL meters based on polarization scrambling have a large measurement uncertainty. The PDL-201 instead searches systematically for the maximum and minimum transmission, so it stays accurate for devices with low PDL and for devices with high PDL. Instruments based on the Mueller Matrix method need wavelength calibration. The PDL-201 covers 1260 to 1620 nm without it.

The instrument has USB, Ethernet, RS-232 and GPIB interfaces for PC control, an OLED display and an analog PDL output port. It is fast and can be controlled remotely, so it can run with tunable lasers in automatic or semi-automatic test stations. It can also measure wavelength-dependent PDL of passive devices, including DWDM and fiber sensor components, in manufacturing and in the laboratory.

Key Features

  • 30 ms measurement speed
  • Wide wavelength range
  • High PDL accuracy
  • Analog PDL output
  • Bright OLED display
  • Power meter function
  • Large measurement dynamic range
  • Max-min search method compliant with TIA/EIA-455-198
  • USB, Ethernet, RS-232 and GPIB interfaces for PC control

Specifications

Measurement
Wavelength Range 1260 to 1620 nm
Resolution 0.01 dB
Web page prints "0.01 db"
PDL Accuracy ± (0.01 + 5% of PDL) (dB)
For 10 sample average; at 23 ± 5°C; accurate PDL measurement also depends on test setup
PDL Repeatability ± (0.005 + 2% of PDL) (dB)
For 10 sample average
PDL Dynamic Range 0 to 45 dB
For input power ≥0 dBm
IL Accuracy ± (0.01 + 5% of IL) (dB)
For 10 sample average; in power meter measurement mode with user-defined reference
IL Repeatability ± (0.005 dB + 2% of IL) (dB)
For 10 sample average
IL Dynamic Range 0 to 45 dB
For input power ≥0 dBm
Measurement Speed 30 ms/measurement
For input > -30 dBm
Optical Power
Optical Power Range -40 to +6 dBm
At DUT output port
Optical Power Accuracy ±0.25 dB
Wavelength Calibration for Power Measurement 1260 to 1360 nm and 1440 to 1620 nm
Interfaces
Communication Interfaces USB, Ethernet, RS-232, and GPIB
Front Panel Display OLED graphic display
Optical Connector Type Light source and DUT input: APC; DUT output: Free space adapter
Connector type options listed: FC, SC
Analog Output 0 to 4V PDL monitor voltage
0 to 3.5V linear with PDL, 4V indicates power low
General
Operating Temperature 0 to 50 °C
Power Supply 100 – 240 VAC, 50 – 60 Hz
Dimensions 2U, 19" half rack width 14" (L) x 8.5" (W) x 3.5" (H)

Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.

Applications

  • PIC and silicon photonics testing
  • Manufacturing test of fiber optic assemblies
  • Characterization of passive components
  • DWDM device characterization
  • PDL vs. wavelength measurement
  • Fiber sensor component characterization

Documents & Resources