MA24106A USB Power Sensor, True-RMS Average Power, 50 MHz to 6 GHz
True-RMS USB power sensor that measures average power from 50 MHz to 6 GHz over a 63 dB dynamic range without a separate power meter.
- Frequency Range
- 50 MHz to 6 GHz
- Dynamic Range
- -40 dBm to +23 dBm
- Input Return Loss
- > 26 dB (50 MHz to < 2 GHz); > 20 dB (2 GHz to 6 GHz)
- SWR
- <1.1 (50 MHz to <2 GHz); <1.2 (2 GHz to 6 GHz)
Overview
The Anritsu MA24106A is a USB power sensor that replaces a traditional power meter. It is a standalone instrument that communicates with a PC over USB. Its measurement capability mimics a thermal (thermo-electric) power sensor while offering a wider dynamic range.
The sensor measures true RMS power regardless of the type or bandwidth of the input signal, which makes it suitable for average power measurements of CW, multi-tone and modulated RF waveforms such as 3G, 4G and OFDM. A micro-controller, signal conditioning circuitry, ADC and power supply are built into the sensor, so it functions as a complete miniature power meter. Data display and acquisition take place on the PC.
Software and compatibility
The Anritsu Power Meter / PowerXpert application for Microsoft Windows PCs is supplied as standard and is used to control the sensor and analyze data. The MA24106A is also compatible with Anritsu MS272xB, MT8222A, MS202xA, MS203xA and MS271xB handheld instruments; in some cases the high accuracy power meter software option (Option 19) is required, and the handheld firmware must be at the level listed by Anritsu.
Calibration
On-board frequency and temperature corrections together with the integrated signal acquisition hardware eliminate the need for a reference calibration, simplifying the measurement process and reducing test time in production. Customers with a suitably equipped calibration laboratory can self-calibrate the sensor using the optional CalXpert software utility. The recommended calibration cycle is 12 months.
Key Features
- True RMS measurements over a 63 dB dynamic range for accurate CW and modulated power measurements
- Suited to installation and maintenance of base stations and testing of 3G and 4G devices, cell phones and general purpose RF devices
- High damage power levels and ESD protection circuitry for ruggedness and reliability
- Low power consumption (100 mA typ) extends laptop battery life
- Eliminates the need for a reference calibrator, reducing test time and handling in production
- Light weight, economical and easy to use with a desktop or laptop PC
- One year calibration cycle and worldwide calibration and service centers
- Compatible with most Anritsu RF and microwave handheld analyzers (Option 19 required in some cases)
- Complimentary PowerXpert PC application for sensor control and data analysis
- Self-calibration possible with the optional CalXpert software utility
Specifications
| Frequency Range | 50 MHz to 6 GHz |
|---|---|
| Dynamic Range | -40 dBm to +23 dBm |
| Input Return Loss | > 26 dB (50 MHz to < 2 GHz); > 20 dB (2 GHz to 6 GHz) |
| SWR | <1.1 (50 MHz to <2 GHz); <1.2 (2 GHz to 6 GHz) |
| Measurement Ranges | Range 1, -40 dBm to -5 dBm; Range 2, -5 dBm to +23 dBm |
| Signal Channel Bandwidth |
100 Hz
typical
|
| RF Connector | N (m) |
| Linearity |
± 0.13 dB (power level < +18 dBm); ± 0.18 dB (power level ≥ +18 dBm)
Web page states Sensor Linearity <3%
|
|---|---|
| Calibration Factor |
± 0.035 dB
Expanded uncertainty with K=2 for absolute power measurements on CW signal at 0 dBm calibration level from 50 MHz to 6 GHz
|
| Noise |
< 2.5 nW (-40 dBm to -5 dBm); < 0.6 µW (-5 dBm to +23 dBm)
Expanded uncertainty with K=2 after zero operation, 128 averages for 5 minutes. In high aperture time mode noise is 1.3 nW and 0.3 µW in Range 1 and Range 2 respectively
|
| Zero Set | < 10 nW (-40 dBm to -5 dBm); < 1.7 µW (-5 dBm to +23 dBm) |
| Zero Drift |
< 3.0 nW (-40 dBm to -5 dBm); < 0.5 µW (-5 dBm to +23 dBm)
After one hour warm-up and zero operation. Measured with 128 averages for one hour keeping the temperature within ± 1 ºC
|
| Temperature Compensation |
± 0.06 dB (at 0 °C to 50 °C)
Measurement error with reference to a CW signal of equal power and frequency at 25 ºC
|
| Effect of Digital Modulation |
± 0.02 dB (power level < +18 dBm); ± 0.10 dB (power level ≥ +18 dBm)
Measurement error with reference to a CW signal of equal power and frequency at 25 ºC
|
| Measurand | True-RMS / Average power |
|---|---|
| Measurement Resolution | 0.01 dB |
| Offset Range | ± 100 dB |
| Averaging Range | 1 to 256 |
| Measurement Speed |
10 measurements per second
typical; one measurement per second, typical in high aperture time mode
|
| Range | Auto-ranging between Range 1 and Range 2 |
| Interface | USB 2.0 |
| Host Operating System | Microsoft Windows 7, Windows Vista, Windows XP, Windows 2000, Windows 8, and Windows 10 |
|---|---|
| Micro Processor |
Intel Pentium III with 1 GB RAM or Intel Pentium IV with 512 MB RAM, or equivalent
Intel Pentium IV with 1 GB RAM recommended
|
| Supporting Software | Microsoft .NET 3.5 |
| Disc Space | 100 MB hard-disk free space |
| Display | Display resolution 1024 × 768 |
| USB Port | USB 2.0 full-speed (compatible with USB 1.0 and 1.1) interface |
| Current (via host USB Mini-B) | 100 mA (typical) at 5 V (150 mA maximum) |
|---|---|
| Maximum DC Voltage at RF Port | ± 25 V |
| Maximum CW Power | +33 dBm |
| Size |
60.4 mm x 22.2 mm x 84.2 mm typical (2.37 in x 0.87 in x 3.31 in)
Not including N connector
|
| Weight | 180 grams typical (6.4 oz) |
| Calibration Cycle |
12 months
Recommended
|
| Warranty | 1 Year |
| European Union EMC | 2014/30/EU, EN 61326:2013, CISPR 11/EN 55011, IEC/EN 61000-4-2/3/4/5/6/8/11 |
|---|---|
| Low Voltage Directive | 2014/35/EU |
| Safety | EN 61010-1:2010 |
| RoHS Directive | 2011/65/EU |
| Australia and New Zealand | RCM AS/NZS 4417:2012 |
| South Korea | KCC-REM-A21-0004 |
| Operating Temperature |
0 °C to 50 °C
Tests were performed per MIL-PRF-28800F (Class 3)
|
|---|---|
| Storage Temperature | -51 °C to 71 °C |
| Maximum Relative Humidity |
45% relative humidity at 55 ºC; 75% relative humidity at 40 ºC; 95% relative humidity at 30 ºC
non-condensing
|
| Mechanical Shock | 30 g half-sine, 11 ms duration |
| Vibration | Sinusoidal: 5-55 Hz, 3 g max; Random: 10-500 Hz, power spectral density 0.03 g²/Hz |
|
MS2721B
MS2721B
|
> V1.24 |
|---|---|
|
MS2723B, MS2724B
MS2723B, MS2724B
|
> V1.21 |
|
MT8222A
MT8222A
|
> V1.44 |
|
MS2024A, MS2026A
MS2024A, MS2026A
|
> V1.55 |
|
MS2034A, MS2036A
MS2034A, MS2036A
|
> V1.11 |
|
MS2717B
MS2717B
|
> V1.05 |
|
MS2718B, MS2719B
MS2718B, MS2719B
|
> V1.04 |
Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.
Applications
- Base station installation and maintenance
- 3G and 4G device testing
- Cell phone testing
- General purpose RF device testing
- Production test
- Average power measurement of CW, multi-tone and OFDM signals
Included
- 2000-1566-R 1.8 meters USB A to Mini-B cable
- 10585-00021 Quick Start Guide
- PowerXpert PC application
Options & Accessories
| Option | Description |
|---|---|
| MA24106A-097 | Option 97, Accredited Calibration to ISO17025 and ANSI/NCSL Z540 — Test report and uncertainty data included. |
| MA24106A-098 | Option 98, Standard calibration to ISO17025 and ANSI/NCSL Z540 |
| MA24106A-099 | Option 99, Premium calibration to ISO17025 and ANSI/NCSL Z540 — Test report and uncertainty data included. |
| 2000-1593-R | 3 meters USB A to Mini-B cable — Optional accessory |
| 2000-1594-R | 5 meters USB A to Mini-B cable — Optional accessory |
| 01-200 | Calibrated torque wrench for N connector — Optional accessory |
| 01-204 | Calibrated torque wrench for K and V connectors — Optional accessory |
| 3-1010-123 | Power attenuator, N(m) to N(f), DC to 8.5 GHz, 30 dB, 50 W, 50 Ω — Recommended for cable and antenna line sweep applications only. Not recommended for low PIM. |
| 3-1010-124 | Power attenuator, N(m) to N(f), DC to 8.5 GHz, 40 dB, 100 W, 50 Ω — Recommended for cable and antenna line sweep applications only. Not recommended for low PIM. |
| 3-1010-122 | Power attenuator, N(m) to N(f), DC to 12.4 GHz, 20 dB, 5 W, 50 Ω — Recommended for cable and antenna line sweep applications only. Not recommended for low PIM. |
| 42N50-20 | Power attenuator, N(m) to N(f), DC to 18 GHz, 20 dB, 5 W, 50 Ω — Recommended for cable and antenna line sweep applications only. Not recommended for low PIM. |
| 42N50-30 | Power attenuator, N(m) to N(f), DC to 18 GHz, 30 dB, 50 W, 50 Ω — Recommended for cable and antenna line sweep applications only. Not recommended for low PIM. |
| 510-90 | Precision coaxial adapter, N(m) to 7/16 DIN(f), DC to 3.3 GHz |
| 510-91 | Precision coaxial adapter, N(f) to 7/16 DIN(f), DC to 3.3 GHz |
| 510-92 | Precision coaxial adapter, N(m) to 7/16 DIN(m), DC to 3.3 GHz |
| 510-93 | Precision coaxial adapter, N(f) to 7/16 DIN(m), DC to 3.3 GHz |
| 33NFNF50B | Precision coaxial adapter, N(f) to N(f), DC to 18 GHz |
| 33NNF50B | Precision coaxial adapter, N(m) to N(f), DC to 18 GHz |
| 33NN50B | Precision coaxial adapter, N(m) to N(m), DC to 18 GHz |
| 34AN50 | Precision coaxial adapter, GPC-7 to N(m), DC to 18 GHz |
| 34ANF50 | Precision coaxial adapter, GPC-7 to N(f), DC to 18 GHz |
| 34NFK50 | Precision coaxial adapter, N(f) to K(m), DC to 18 GHz |
| 34NFKF50 | Precision coaxial adapter, N(f) to K(f), DC to 18 GHz |
| 34NK50 | Precision coaxial adapter, N(m) to K(m), DC to 18 GHz |
| 34NKF50 | Precision coaxial adapter, N(m) to K(f), DC to 18 GHz |
| 2300-528 | MA24106A CalXpert — Sensor calibration utility |