Anritsu Power Master MA24507A and MA24510A mmWave power analyzers
MA24510A

Power Master MA24510A 9 kHz to 110 GHz mmWave Power Analyzer

USB-powered, frequency-selectable mmWave power analyzer covering 9 kHz to 110 GHz with a measurement floor as low as -90 dBm.

Frequency Range
9 kHz to 110 GHz
RF Connector
W1(m) Connector (1.0 mm)
Internal Reference Accuracy
±0.2 ppm (0 °C to 50 °C)
Internal Reference Aging
±1.0 ppm/year aging
Datasheet

Overview

The Anritsu Power Master MA24510A is a USB-powered mmWave power analyzer that makes simple, numeric, frequency-selective average power measurements from 9 kHz to 110 GHz. Its measurement range reaches down to -90 dBm, and it can measure the amplitude of up to six signals at once. The RF input is a W1 male (1.0 mm) connector.

The Power Master uses a receiver-based architecture rather than the thermal components found in many mmWave sensors. This shortens measurement times and makes the measurement frequency-selective, so power is measured only in the frequency band of interest. The instrument is slightly larger than a smartphone and takes up little bench space.

PowerXpert software

The analyzer is controlled from the PowerXpert PC application, which handles the general settings (measurement mode, center frequency, span and resolution). In Power Hunter mode, the user sets a frequency range and the Power Master finds the six highest CW amplitudes and their frequencies. Channel Monitor mode tracks the CW amplitude or channel power of up to six channels, each up to 20 MHz wide, at the same time.

Use cases

The Power Master can be used for stand-alone measurements or as part of a system for calibrating vector network analyzers. Because it can measure very low power levels, it suits over-the-air testing of mmWave signals with high propagation loss. Mounting holes let it attach directly to probes for over-the-air or on-wafer testing.

Key Features

  • Measures signals as low as –90 dBm
  • Frequency-selective measurement from 9 kHz to 110 GHz
  • Suited to over-the-air testing of mmWave signals with high propagation loss
  • User settings to control measurement speed and noise floor
  • Channel Monitor mode in PowerXpert for monitoring up to six frequency channels at once
  • Power Hunter mode in PowerXpert for searching up to six signals within a frequency range
  • Mounting holes for direct mounting to probes for over-the-air or on-wafer testing
  • Receiver-based architecture for reduced measurement time
  • Powered over USB in a small form factor

Specifications

Frequency
Frequency Range 9 kHz to 110 GHz
All performance above 70 GHz is typical
RF Connector W1(m) Connector (1.0 mm)
Internal Reference Accuracy ±0.2 ppm (0 °C to 50 °C)
Internal Reference Aging ±1.0 ppm/year aging
Continuous Mode Span 30 kHz to 2 GHz max in Channel Power Measurement; 10 kHz to Full Span in CW Max Measurement
Channel Monitor Mode Span 1 kHz to 20 MHz
Power Measurement
Measurement Range -90 — +10 dBm
Maximum Amplitude (≤ 6.15 GHz) +15 dBm
Characteristic
Maximum Amplitude (> 6.15 GHz) +10 dBm
Characteristic
Damage Level, Continuous +30 dBm CW, +/- 10 VDC max
Range, Lower ≤ –10 dBm
The user defines the operating range; the upper range is recommended for signals above –10 dBm to avoid clipping or saturation
Range, Upper > –10 dBm
Trigger Source Bus, Continuous
Average Noise Floor
Channel Power Measurement, 30 kHz channel span –88 dBm
Measured at 1 GHz center frequency
Channel Power Measurement, 10 MHz channel span –64 dBm
Measured at 1 GHz center frequency
Channel Power Measurement, 1 GHz channel span –40 dBm
Measured at 1 GHz center frequency
CW Max Measurement, High resolution –100 dBm
Measured at 1 GHz center frequency; 3 MHz span
CW Max Measurement, Medium resolution –90 dBm
Measured at 1 GHz center frequency; 3 MHz span
CW Max Measurement, Low resolution –80 dBm
Measured at 1 GHz center frequency; 3 MHz span
Input Match (W1 Connector, typical)
9 kHz to 12.4 GHz VSWR 1.29:1, Return Loss 18 dB
Typical
> 12.4 GHz to 26.5 GHz VSWR 1.67:1, Return Loss 12 dB
Typical
> 26.5 GHz to 40 GHz VSWR 1.67:1, Return Loss 12 dB
Typical
> 40 GHz to 50 GHz VSWR 1.67:1, Return Loss 12 dB
Typical
> 50 GHz to 70 GHz VSWR 2.10:1, Return Loss 9 dB
Typical
> 70 GHz to 110 GHz VSWR 2.10:1, Return Loss 9 dB
Typical
Measurement Speed (readings/s, characteristic)
Channel Power Measurement 7 (300 kHz span), 20 (20 MHz span), 10 (1 GHz span)
Measured at 1 GHz center frequency; no averages
CW Max Measurement (High) 0.8 (300 kHz span), 15 (20 MHz span), 6 (1 GHz span)
Measured at 1 GHz center frequency; no averages
CW Max Measurement (Medium) 4 (300 kHz span), 25 (20 MHz span), 10 (1 GHz span)
Measured at 1 GHz center frequency; no averages
CW Max Measurement (Low) 20 (300 kHz span), 25 (20 MHz span), 10 (1 GHz span)
Measured at 1 GHz center frequency; no averages
Measurement Uncertainty
Amplitude Accuracy, 9 kHz to 644 MHz ±1.3 dB max, ±0.5 dB typ. (20 °C to 30 °C); ±2.0 dB max, ±0.5 dB typ. (0 °C to 50 °C)
20 °C to 30 °C after 30 minute warm-up; 0 °C to 50 °C after 60 minute warm-up. Excludes effects of noise and mismatch uncertainty
Amplitude Accuracy, > 644 MHz to 40 GHz ±1.8 dB max, ±0.5 dB typ. (20 °C to 30 °C); ±3.0 dB max, ±1.0 dB typ. (0 °C to 50 °C)
Excludes effects of noise and mismatch uncertainty
Amplitude Accuracy, > 40 GHz to 70 GHz ±2.0 dB max, ±0.5 dB typ. (20 °C to 30 °C); ±3.0 dB max, ±1.0 dB typ. (0 °C to 50 °C)
Excludes effects of noise and mismatch uncertainty
Amplitude Accuracy, > 70 GHz to 90 GHz ±2.2 dB max, ±0.5 dB typ. (20 °C to 30 °C); ±3.0 dB max, ±1.0 dB typ. (0 °C to 50 °C)
All performance above 70 GHz is typical
Amplitude Accuracy, > 90 GHz to 110 GHz ±2.5 dB max, ±0.5 dB typ. (20 °C to 30 °C); ±3.0 dB max, ±1.0 dB typ. (0 °C to 50 °C)
All performance above 70 GHz is typical
Amplitude Accuracy (summary) ±1.0 dB, typical
Web page summary value; datasheet gives the frequency-dependent table above
Relative Power Accuracy, 9 kHz to < 6.15 GHz ±0.3 dB
Relative Power Accuracy, 6.15 GHz to < 40 GHz ±0.3 dB
Relative Power Accuracy, 40 GHz to ≤ 110 GHz ±0.3 dB
Typical with W1 connector
PowerXpert
Processor and RAM Equivalent to Quad Core i5 fourth generation or higher CPU, 8 GB RAM
PC requirements, PowerXpert version 4.0 or greater
Operating System Microsoft Windows 10, 8.1, or 7; 64-bit
Hard-Disk Free Space 100 MB minimum
Display Resolution 1024 × 768 minimum
Interface USB 3.0
Measurand Channel power, CW peak power
Measurement Resolution 0.01 dB max via PowerXpert, 0.01 dB max via remote command
Offset Correction –100 dB to +150 dB
Available only through the PowerXpert application; no remote command in the analyzer firmware
Units dBm, nW, µW, mW, W
Averaging Manual; Moving; 1 to 1,000 averages
Continuous Mode Center Frequency 9.5 kHz to (Max Freq – 500 Hz)
Continuous Mode Span 30 kHz to 2 GHz (Channel power), 1 kHz to Full span (CW max)
Continuous Mode Resolution High, medium, low
Power Hunter Mode Measurement CW max only
Power Hunter Start Frequency 9 kHz to (Max Freq – 1 kHz)
Power Hunter Stop Frequency 10 kHz to Max Freq
Power Hunter Set Minimum Power Range –130 to 0 dBm
Channel Monitor Measurements Channel power, CW max
Channel Monitor Channel Frequencies (9 kHz + Span/2) to (Max Freq – Span/2)
Channel Monitor Span 1 kHz to 20 MHz
Number of Channels Up to 6
General
RF Connector W1 male (1.0 mm)
Trigger Port Inactive
Reference Port For Service Only
Interface to Host USB 3.0
Current Consumption 900 mA max
Size 155 mm x 84 mm x 27 mm (6.1 in x 3.3 in x 1.1 in)
Weight 282 g (0.62 lb)
Warranty 1 year
Warm-Up Time 30 minutes
Specification conditions unless otherwise stated
Calibration Cycle Anritsu recommended calibration interval is 12 months
Environmental
Operational Requirements Tests were performed per MIL-PRF-28800F (Class 3)
Operating Temperature Range 0 ºC to 50 ºC
Storage Temperature Range –40 ºC to +71 ºC
Relative Humidity (non-condensing) 45 % at 50 ºC, 75 % at 40 ºC, 95 % at 30 ºC
Altitude 4600 m operational max
Shock 30 g half-sine, 11 ms duration
Vibration Sinusoidal: 5 Hz to 55 Hz, 3 g max; Random: 10 Hz to 500 Hz, 2.34 g rms; Power Spectral Density: 0.01 g²/Hz
Regulatory Compliance
European Union EMC 2014/30/EU, EN 61326:2013, CISPR 11/EN 55011, IEC/EN 61000-4-2/3/4/5/6/8/11
Low Voltage Directive 2014/35/EU
Safety EN 61010-1:2010
RoHS Directive 2011/65/EU
Australia and New Zealand RCM AS/NZS 4417:2012
South Korea KCC-REM-A21-0004

Specifications are taken from the manufacturer's documentation. See the datasheet for complete and current values.

Applications

  • Over-the-air mmWave testing
  • On-wafer testing
  • Vector network analyzer calibration systems
  • Stand-alone RF power measurements
  • Antenna quality control testing for consumer electronics
  • Parallel remote monitoring with multiple Power Masters

Included

  • 2000-1859-R USB 3.0 Type C to Type A Cable, 1 m
  • 2000-1605-R BNC(m) to MCX(m) Cable (qty 2)

Options & Accessories

Option Description
MS24510A-098 Option 98: Standard calibration ISO/IEC 17025 and ANSI/NCSL Z540-1 — Printed in the datasheet as MA24507A-098/MS24510A-098
MS24510A-099 Option 99: Premium calibration ISO/IEC 17025 and ANSI/NCSL Z540-1 — Includes test report and uncertainty data. Printed in the datasheet as MA24507A-099/MS24510A-099
41W-3 Precision Fixed Attenuator — Precision, DC to 110 GHz, 1W, 3 dB, W1(m) to W1 (f)
41W-6 Precision Fixed Attenuator — Precision, DC to 110 GHz, 1W, 6 dB, W1(m) to W1 (f)
41W-10 Precision Fixed Attenuator — Precision, DC to 110 GHz, 1W, 10 dB, W1(m) to W1 (f)
W240A Precision Power Divider — Precision, DC to 110 GHz, 3 resistor, W1(f) input, W1(f) output
W241A Precision Power Splitter — Precision, DC to 110 GHz, 2 resistor, W1(m) input, W1(f) output
MN25110A Precision Directional Coupler — Precision, 20 GHz to 110 GHz, W1(f) input, W1(f) output, W1(f) coupled port
34WFV50 Precision Coaxial Adapter — DC to 65 GHz, W1(f) to V(m), 50 Ω
34WFVF50 Precision Coaxial Adapter — DC to 65 GHz, W1(f) to V(f), 50 Ω
34WV50 Precision Coaxial Adapter — DC to 65 GHz, W1(m) to V(m), 50 Ω
33WFWF50 Precision Coaxial Adapter — DC to 110 GHz, W1(f) to W1(f), 50 Ω
33WW50 Precision Coaxial Adapter — W1(m) to W1(m)
33WWF50 Precision Coaxial Adapter — W1(m) to W1(f)
35WR10WF Waveguide to Coaxial Adapter (right angle) — 75 GHz to 110 GHz, WR10 to W1(f)
SC7442 Waveguide to Coaxial Adapter (right angle) — 60 GHz to 90 GHz, WR12 to W1(f)
1091-400-R Waveguide to Coaxial End Launch Adapter (straight through) — 73.8 GHz to 110 GHz, WR10 to W1(f)
1091-401-R Waveguide to Coaxial End Launch Adapter (straight through) — 60.5 GHz to 92.0 GHz, WR12 to W1(f)
1091-402-R Waveguide to Coaxial End Launch Adapter (straight through) — 49.9 GHz to 75.8 GHz, WR15 to W1(f)
2000-1867-R Directional Horn Antenna — 17.6 GHz to 26.7 GHz, WR42, 25 dBi gain
2000-1868-R Directional Horn Antenna — 26.4 GHz to 40.1 GHz, WR28, 25 dBi gain
2000-1869-R Directional Horn Antenna — 33.0 GHz to 50.1 GHz, WR22, 25 dB gain
2000-1870-R Directional Horn Antenna — 39.3 GHz to 59.7 GHz, WR19, 25 dBi gain
2000-1871-R Directional Horn Antenna — 49.9 GHz to 75.8 GHz, WR15, 25 dBi gain
2000-1872-R Directional Horn Antenna — 60.0 GHz to 90.0 GHz, WR12, 25 dBi gain
2000-1873-R Directional Horn Antenna — 75.0 GHz to 110.0 GHz, WR10, 25 dBi gain
2000-1888-R USB 3.0 Powered Extension Cable, 10 m, (32 ft) — Up to two can be used in series for a total length of 20 m